20 of 62
0319 IH-1679
ERROR CODES
ERROR CODES
DESCRIPTION
RESOLUTION
-----
Overload
Remove weight from the scale.
If the problem persists contact Uline Customer
Service.
Err 1
Date Setting Error
Enter date using correct format.
Format: yy:mm:dd
Err 2
Time Setting Error
Enter time using correct format.
Format: hh:mm:ss
Err 4
Zero Setting Error
The scale was outside the normal zero setting
range either when it was turned on or when
the ZERO key was pressed.
Remove weight from the scale and try again.
Use the TARE key to set the display to zero value.
If the problem persists contact Uline Customer
Service.
Err 6
A/D out of range
The values from the A/D converter are outside
the normal range. Remove weight from the
scale if overloaded, make sure fork cover
plate is attached.
Indicates the load cell or the electronics may
be faulty.
If the problem persists contact Uline Customer
Service.
π
CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA
1-800-295-5510
uline.com
Summary of Contents for H-1679
Page 4: ...4 of 62 0319 IH 1679 LCD DISPLAY Low Battery Symbol Net Weight Mode Weighing Units...
Page 24: ...24 of 62 0319 IH 1679 PANTALLA LCD S mbolo de Bater a Baja Modo Peso Neto Unidades de Pesaje...
Page 45: ...45 of 62 0319 IH 1679 AFFICHAGE ACL Indicateur de d charge Mode poids net Unit s de pesage...