H-2289, H-2290, H-2291

OHaus DefenDer™  

5000 DIGITaL scaLe

1-800-295-5510

uline.com

Overview Of COntrOls

COntrOl Panel

#

DesCriPtiOn

1

Under LED

2

Accept LED

3

Over LED

4

Capacity Label Window

5

TARE Menu Button

6

FUNCTION Mode Button

7

PRINT Units Button

8

ON/ZERO Off Button

ButtOn

Primary funCtiOn 

(Short Press)

On/ZerO

Turns the indicator ON
If indicator is ON, sets 

zero.

Print

Sends the current value to 

the COM port if AUTOPRINT 

is set to off.

funCtiOn

Initiates an application mode.
Temporarily displays the active 

mode's reference data.
In Weigh mode, temporarily 

displays 10x expanded resolution

tare

Performs a tare 

operation.

seCOnDary funCtiOn 

(Long Press)

Off

Turns the indicator off.

units

Changes the weighing 

unit.

mODe

Allows changing the application 

mode. 
Press and hold allows scrolling 

through modes.

menu

Enter the User Menu.

menu funCtiOn 

(Short Press)

yes

Accepts the current 

setting on the display. 

nO

Advances to the next 

menu or menu item. 
Rejects the current setting 

on the display and 

advances to the next 

available setting. 
Increments the value. 

BaCK

Moves back to previous menu 

item. 
Decrements the value. 

eXit

Exits the User Menu. 
Aborts the calibration 

in progress. 

COntrOl funCtiOns

COntrOl Panel Parts

On/Zero

Off

Yes

Print

Units

No

function

Mode

Back

tare

Menu

Exit

lb

Print
Units

No

Function

Mode

Back

Tare

Menu

Exit

On/Zero

Off

Yes

Under

Over

Accept

1 2 3

4

5

6

7

8

Summary of Contents for H-2289

Page 1: ...xpanded resolution TARE Performs a tare operation Secondary Function Long Press OFF Turns the indicator off UNITS Changes the weighing unit MODE Allows changing the application mode Press and hold all...

Page 2: ...or End menu items pressing the YES button advances to the next menu while pressing the NO button returns to the top of the current menu CALIBRATION MENU Three calibration processes are available Zero...

Page 3: ...d by the user during the calibration procedure 1 When LINEAr is displayed press the YES button to initiate linearity calibration 2 The display flashes the full calibration point and calibration unit P...

Page 4: ...y The scale must be stable to accept zero operation manual tare When weighing an item that must be held in a container taring stores the container weight in memory 1 Place the empty container on the s...

Page 5: ...tton to clear the stored APW recalling a stored apw 1 Press the NO button to recall the existing APW NOTE Press the FUNCTION MODE button to temporarily display the APW value 2 The display shows the sa...

Page 6: ...Refer to Menu Navigation to enter settings NOTE The minus sign is used together with the first digit to show a negative value Editing the OVER setting 1 The display shows SEt HI Press the YES button...

Page 7: ...l pertinent legal requirements are met Settings Before verification and sealing perform the following steps 1 Verify that the menu settings meet the local weights and measures regulations 2 Perform a...

Page 8: ...Error 8 2 Weight reading below Power On Zero limit Add load to scale Recalibrate scale Error 8 3 Weight reading exceeds Overload limit Reduce load on scale Error 8 4 Weight reading below Underload lim...