background image

PAGE 6 OF 8 

1214 IH-2289

OPeratiOn COntinueD

referenCe weight COntinueD

ESTABLISHING A NEW REFERENCE WEIGHT

The display shows Put.rEF %.

1.  To establish the reference weight, 

place the sample on the scale 

and press the FUNCTION MODE 

button to capture the weight. rEF.

Wt is displayed shortly followed by 

the value with the current unit of 

measure.

BEGIN PERCENT WEIGHING

1.  Place the sample on the scale 

and read the percent value. If a 

container is used, be sure to tare 

the empty container first.

CheCK weighing

Use this mode to determine if 

the weight of a sample is within 

prescribed limits.

CheCK weighing limits

When the FUNCTION MODE button is 

released, CLr.rEF is displayed.

CLEARING STORED CHECk WEIGHING LIMITS

1.  Press the YES button to clear the stored limits.

RECALLING STORED CHECk WEIGHING LIMITS

1.  Press the NO button to recall the stored limits.

nOte: Press the 

FUNCTION 

MODE 

button to temporarily 

display the under and over 

limit values.

EDITING THE UNDER SETTING

1.  The display shows SEt.LO. Press the 

YES button to edit setting.

Settings:

-999950 to 999950

Refer to Menu Navigation 

to enter settings.

nOte: the minus sign is used 
together with the first digit to 
show a negative value.

EDITING THE OVER SETTING

1.  The display shows SEt.HI. Press the 

YES button to edit setting.

Settings:

-999950 to 999950

Refer to Menu Navigation 

to enter settings.

Begin CheCK weighing

The appropriate under, accept or over 

LED lights indicate check weigh status.

1.  Place a sample on the scale and 

read the weight.

2.  For loads less than the under limit, 

the yellow under LED is lit.

3.  For loads greater than the under 

limit and less than the over limit, the 

green accept LED is lit.

4.  For loads greater than the over 

limit, the red over LED is lit.

DynamiC weighing

Use this mode to weigh moving or 

oversized objects. The weight is held 

on the display until reset. Manual, 

semi-automatic and automatic start/

stop methods are available.

Begin DynamiC weighing

1.  When the display shows rEAdY, 

place the object on the scale.

2.  If the manual mode is in use, 

press the FUNCTION MODE button 

to start measurement. If the semi-automatic or 

automatic mode is in use, measurement is started 

automatically.

CLr.rEF 

kg

110.000 

kg

120.000 

kg

SEt.LO 

kg

SEt.HI 

 kg

 kg

CHEC

rEAdY

. O.000

 

kg

. O.123

 

kg

. 1

 

1

 

1.234 

kg

.

 

122.345

  kg

dYNA     

kg 

. 100.00

 

%

  12.345 

kg

Put.rEF 

%

rEF.    t

 kg

 kg

to

Summary of Contents for H-2289

Page 1: ...xpanded resolution TARE Performs a tare operation Secondary Function Long Press OFF Turns the indicator off UNITS Changes the weighing unit MODE Allows changing the application mode Press and hold all...

Page 2: ...or End menu items pressing the YES button advances to the next menu while pressing the NO button returns to the top of the current menu CALIBRATION MENU Three calibration processes are available Zero...

Page 3: ...d by the user during the calibration procedure 1 When LINEAr is displayed press the YES button to initiate linearity calibration 2 The display flashes the full calibration point and calibration unit P...

Page 4: ...y The scale must be stable to accept zero operation manual tare When weighing an item that must be held in a container taring stores the container weight in memory 1 Place the empty container on the s...

Page 5: ...tton to clear the stored APW recalling a stored apw 1 Press the NO button to recall the existing APW NOTE Press the FUNCTION MODE button to temporarily display the APW value 2 The display shows the sa...

Page 6: ...Refer to Menu Navigation to enter settings NOTE The minus sign is used together with the first digit to show a negative value Editing the OVER setting 1 The display shows SEt HI Press the YES button...

Page 7: ...l pertinent legal requirements are met Settings Before verification and sealing perform the following steps 1 Verify that the menu settings meet the local weights and measures regulations 2 Perform a...

Page 8: ...Error 8 2 Weight reading below Power On Zero limit Add load to scale Recalibrate scale Error 8 3 Weight reading exceeds Overload limit Reduce load on scale Error 8 4 Weight reading below Underload lim...

Reviews: