1214 IH-2289
OPeratiOn COntinueD
Begin DynamiC weighing COntinueD
nOte: when using manual
mode, it is not necessary
for the display to be at zero
gross or net. when using
semi-automatic or automatic
mode, the display must be
at zero gross or net before
placing the object on the
scale. the example is for a
setting of 5 seconds. During
the averaging period, the
countdown timer decreases
in one second increments.
nOte: if set 0 was selected in
the Dynamic menu item, the
countodwn timer is not displayed.
3. When the countdown has
completed, the readings are
averaged and held on the display.
The averaged weight is displayed
until rest.
4. If the manual or semiautomatic
mode is in use, reset the countdown
timer by pressing the FUNCTION
MODE button. Then the display
shows rEAdY.
5. If the automatic mode is in use,
the held reading is shown on
the display for 10 seconds after
the object is removed to within 5
divisions of zero. Then the display
shows rEAdY.
6. The scale is now ready to accept a new object.
legal fOr traDe
When the indicator is used in trade or a legally
controlled application, it must be set up, verified and
sealed in accordance with local weights and measures
regulations. It is the responsibility of the purchaser to
ensure that all pertinent legal requirements are met.
SETTINGS
Before verification and sealing, perform the following steps:
1. Verify that the menu settings meet the local weights
and measures regulations.
2. Perform a calibration.
3. Set Legal for Trade to ON in the Setup menu.
4. Without exiting the menu, turn the indicator off.
5. Disconnect power from the indicator and open the
housing.
6. Set the security switch to ON.
7. Close the housing.
8. Reconnect power and turn the indicator on.
nOte: when legal for trade is set to On and the
security switch is set to On, the following menu
settings cannot be changed: Zero Calibration,
span Calibration, linearity Calibration, geO,
range, Capacity, graduation, Power On unit,
Zero range, auto tare, retain weight, legal for
trade, stable range, auto Zero tracking, gross
indicator, modes, units, stable Only.
verifiCatiOn
The local weights and measures official or authorized
service agent must perform the verification procedure.
sealing
The local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings. (See Figures 2 and 3)
LOAD CELL SEALING COLLAR
When the scale base is attached to the indicator using
a connector, it is necessary to seal the load cell cable
to the indicator in some jurisdictions. The load cell
sealing collar is available as an accessory.
(See Figure 4)
Wire Seal
Figure 2
Paper Seal
Figure 3
Collar Seal
Collar Seal
Figure 4
. 1.234
kg
~
rEAdY
5 SEC
1 SEC
•
•
•
. 1.234
kg
~