1214 IH-2289

OPeratiOn COntinueD

Begin DynamiC weighing COntinueD

nOte: when using manual 
mode, it is not necessary 
for the display to be at zero 
gross or net. when using 
semi-automatic or automatic 
mode, the display must be 
at zero gross or net before 
placing the object on the 
scale. the example is for a 
setting of 5 seconds. During 
the averaging period, the 

countdown timer decreases 

in one second increments.

nOte: if set 0 was selected in  
the Dynamic menu item, the  

countodwn timer is not displayed.

3.  When the countdown has 

completed, the readings are 

averaged and held on the display. 

The averaged weight is displayed 

until rest.

4.  If the manual or semiautomatic 

mode is in use, reset the countdown 

timer by pressing the FUNCTION 
MODE 
button. Then the display 

shows rEAdY.

5.  If the automatic mode is in use, 

the held reading is shown on 

the display for 10 seconds after 

the object is removed to within 5 

divisions of zero. Then the display  

shows rEAdY.

6.  The scale is now ready to accept a new object.

legal fOr traDe

When the indicator is used in trade or a legally 

controlled application, it must be set up, verified and 

sealed in accordance with local weights and measures 

regulations. It is the responsibility of the purchaser to 

ensure that all pertinent legal requirements are met.

SETTINGS

Before verification and sealing, perform the following steps:

1.  Verify that the menu settings meet the local weights 

and measures regulations.

2.  Perform a calibration.
3.  Set Legal for Trade to ON in the Setup menu.

4.  Without exiting the menu, turn the indicator off.

5.  Disconnect power from the indicator and open the 

housing.

6.  Set the security switch to ON. 

7.  Close the housing.

8.  Reconnect power and turn the indicator on.

nOte: when legal for trade is set to On and the 

security switch is set to On, the following menu 
settings cannot be changed: Zero Calibration, 
span Calibration, linearity Calibration, geO, 
range, Capacity, graduation, Power On unit, 
Zero range, auto tare, retain weight, legal for 
trade, stable range, auto Zero tracking, gross 
indicator, modes, units, stable Only.

verifiCatiOn

The local weights and measures official or authorized 

service agent must perform the verification procedure.

sealing

The local weights and measures official or authorized 

service agent must apply a security seal to prevent 

tampering with the settings. (See Figures 2 and 3)

LOAD CELL SEALING COLLAR

When the scale base is attached to the indicator using 

a connector, it is necessary to seal the load cell cable 

to the indicator in some jurisdictions. The load cell 

sealing collar is available as an accessory.  

(See Figure 4)

Wire Seal

Figure 2

Paper Seal

Figure 3

Collar Seal

Collar Seal

Figure 4

.    1.234 

kg

~

rEAdY

5  SEC

1  SEC

.    1.234 

kg

~

Summary of Contents for H-2289

Page 1: ...xpanded resolution TARE Performs a tare operation Secondary Function Long Press OFF Turns the indicator off UNITS Changes the weighing unit MODE Allows changing the application mode Press and hold all...

Page 2: ...or End menu items pressing the YES button advances to the next menu while pressing the NO button returns to the top of the current menu CALIBRATION MENU Three calibration processes are available Zero...

Page 3: ...d by the user during the calibration procedure 1 When LINEAr is displayed press the YES button to initiate linearity calibration 2 The display flashes the full calibration point and calibration unit P...

Page 4: ...y The scale must be stable to accept zero operation manual tare When weighing an item that must be held in a container taring stores the container weight in memory 1 Place the empty container on the s...

Page 5: ...tton to clear the stored APW recalling a stored apw 1 Press the NO button to recall the existing APW NOTE Press the FUNCTION MODE button to temporarily display the APW value 2 The display shows the sa...

Page 6: ...Refer to Menu Navigation to enter settings NOTE The minus sign is used together with the first digit to show a negative value Editing the OVER setting 1 The display shows SEt HI Press the YES button...

Page 7: ...l pertinent legal requirements are met Settings Before verification and sealing perform the following steps 1 Verify that the menu settings meet the local weights and measures regulations 2 Perform a...

Page 8: ...Error 8 2 Weight reading below Power On Zero limit Add load to scale Recalibrate scale Error 8 3 Weight reading exceeds Overload limit Reduce load on scale Error 8 4 Weight reading below Underload lim...