0321 IH-7208
SCALE CALIBRATION/OPERATION
1. Turn power off. Remove 'X1' plate from bottom of
display to expose calibration switch. (See Figure 2)
2. Slide calibration switch towards off/on switch to enter
calibration mode. (See Figure 3)
3. Turn power on. Display should read 'F1'.
4. TO SET 5,000 LB. SCALE:
• Press ZERO and verify that screen shows '5,000
'
.
Press SET. Press UNITS. (See Figure 4)
NOTE: If screen does not show '5,000,' press
UNITS or ZERO buttons.
• Press PRINT or TARE and scroll to F9. Press ZERO.
(See Figure 5)
• Set value to '1'. Press NET/GROSS to save.
Press UNITS to exit. (See Figure 6)
• Press PRINT to scroll to 'F10'. Press ZERO and verify that
value is '0'. Press NET/GROSS to save. Press UNITS to exit.
(See Figure 7)
NOTE: If screen does not show '5,000' press
UNITS or ZERO buttons.
5. TO STABILIZE SCALE:
• Remove any items from scale.
• Use PRINT or TARE to scroll to
'
F16
'
.
• Press ZERO.
• When stabilized, green light will appear.
(See Figure 7)
• Press ZERO. Press NET/GROSS. Screen will return to 'F16'.
6. TO CALIBRATE SPAN:
• Using PRINT or TARE, scroll to 'F17'. Press ZERO to enter
Span Calibration Menu.
• Use PRINT, TARE and ZERO to enter test weight (must be
at least 10% of scale capacity).
NOTE: PRINT and TARE move digit left or right.
UNITS and ZERO changes flashing digit.
• Place test weight on scale and allow unit to stabilize for
at least 30 seconds. Press NET/GROSS to return to 'F17'.
• Do not remove weight. Slide calibration switch on
back to original position. Meter will count down and
read test weight on screen.
• Reinstall 'X1' plate.
• Scale is now calibrated and ready for use.
Calibration Switch
Figure 3
GROSS
NET
SET
TARE
ZERO
UNITS
Figure 4
GROSS
NET
SET
TARE
ZERO
UNITS
Figure 5
GROSS
NET
SET
TARE
ZERO
UNITS
Figure 6
GROSS
NET
SET
TARE
ZERO
UNITS
Green Light
Figure 7
X1 Plate
Figure 2
CALIBRACIÓN/FUNCIONAMIENTO DE LA BÁSCULA
SCALE CALIBRATION/OPERATION