ERROR MESSAGES: 

If the scale encounters an error, it will display a message alerting the operator.  A 
description follows: 
 

Code Mode 

Meaning/Possible 

Solution 

 
 

Normal Operating Mode 

Gross Overload.  A weight 
greater than the scale capacity 
has been put on the scale.  
Remove the weight from the 
platter or try recalibrating the 
scale.  Otherwise, check for a 
bad load cell connection or 
possible load cell damage due to 
overloading 

Err 0 

Span Calibration 
Mode (F17) 

Keyed-in weight value is larger 
than the full scale capacity.  Use 
a smaller test weight or check 
keyed-in value. 

ERR 1 

Span Calibration 
Mode (F17) 

Keyed-in weight value is less 
than 1% of full scale capacity.  
Use a larger test weight or check 
keyed-in value. 

ERR 2 

Span Calibration 
Mode (F17) 

There is not enough load cell 
signal to produce the internal 
counts necessary to properly 
calibrate the scale.  First check 
all load connections.  Use F16 
mode to view internal counts.  

ERR 3 

All Modes 

Non-volatile memory read error.  
One or more setup parameters 
have been lost. 

ERR 4 

All Modes  

Non-volatile memory write error.  
Indicator needs service. 

ERR 5 

Key-in Span Calibration Mode 
(F20) 

You have attempted to enter a 
zero value for C1.  Enter a 
known calibration value greater 
than zero. 

ERR 7 

Initialization 

No reading from the ADC.  Make 
sure there is a load cell 
connected to the indicator at 
start-up 

 

 