ERROR MESSAGES:
If the scale encounters an error, it will display a message alerting the operator. A
description follows:
Code Mode
Meaning/Possible
Solution
Normal Operating Mode
Gross Overload. A weight
greater than the scale capacity
has been put on the scale.
Remove the weight from the
platter or try recalibrating the
scale. Otherwise, check for a
bad load cell connection or
possible load cell damage due to
overloading
Err 0
Span Calibration
Mode (F17)
Keyed-in weight value is larger
than the full scale capacity. Use
a smaller test weight or check
keyed-in value.
ERR 1
Span Calibration
Mode (F17)
Keyed-in weight value is less
than 1% of full scale capacity.
Use a larger test weight or check
keyed-in value.
ERR 2
Span Calibration
Mode (F17)
There is not enough load cell
signal to produce the internal
counts necessary to properly
calibrate the scale. First check
all load connections. Use F16
mode to view internal counts.
ERR 3
All Modes
Non-volatile memory read error.
One or more setup parameters
have been lost.
ERR 4
All Modes
Non-volatile memory write error.
Indicator needs service.
ERR 5
Key-in Span Calibration Mode
(F20)
You have attempted to enter a
zero value for C1. Enter a
known calibration value greater
than zero.
ERR 7
Initialization
No reading from the ADC. Make
sure there is a load cell
connected to the indicator at
start-up