PAGE 13 OF 54
0521 IH-5276
OPERATION CONTINUED
Confirm the default values displayed (Water Temp.)
are correct.
To edit the default values, touch
ITEM SETTINGS
. The
settings screen appears. (See Figure 47)
The balance calculates water density based on the
water temperature value entered (look-up table).
Measure the actual water temperature using a precision
thermometer. To adjust the water temperature or oil
density values, touch the
WATER TEMP.
or
OIL DENSITY
button.
Numeric input windows appear. (See Figures 48-49)
Key in the desired value, then press
SAVE
.
The display returns to the previous screen with the new
value highlighted. To return to the density determination
home screen, touch
EXIT
.
Press
BEGIN DENSITY CALCULATION
.
STEP 1 OF 3 – WEIGH THE UN-OILED SAMPLE
IN AIR.
Follow the screen instructions then press
ACCEPT
to store
the dry sample weight (in air). (See Figure 50)
STEP 2 OF 3 – WEIGH THE OILED SAMPLE IN AIR.
Follow the screen instructions then press
ACCEPT
to store
the sample weight (oiled). (See Figure 51)
STEP 3 OF 3 – WEIGH THE OILED SAMPLE
SUBMERGED IN LIQUID.
Follow the screen instructions then press
ACCEPT
to
store the oiled sample weight (submerged in liquid).
(See Figure 52)
Figure 47
Figure 48
Figure 49
Density Determination
Porous Material
Water Temp.
Auxilliary Liquid
Density Type
Oil Density
Print Settings
Exit
Back
Oil Density
0.8000 g/cm3
1
2
3
4
5
6
7
8
9
Clear
0
•
Exit
Save
Water Temp.
20.0 ºC
1
2
3
4
5
6
7
8
9
Clear
0
•
Exit
Save
Figure 50
249.94
Density Determination
Menu
Stop
Accept
Item
Settings
------
------
Dry Weight:
Olied Weight:
------
20.0ºC
Liquid Weight:
Water Temp.:
g
Max 4200 g
d=0.01 g
Weigh sample in air. Press Accept.
Figure 51
249.94
Density Determination
Menu
Stop
Accept
Item
Settings
249.94g
------
Dry Weight:
Olied Weight:
------
20.0ºC
Liquid Weight:
Water Temp.:
g
Max 4200 g
d=0.01 g
Weigh oiled sample in air. Press Accept.
Figure 52
78.37
Density Determination
Menu
Stop
Accept
Item
Settings
249.94 g
249.94 g
Dry Weight:
Olied Weight:
------
20.0ºC
Liquid Weight:
Water Temp.:
g
Max 4200 g
d=0.01 g
Weigh oiled sample in liquid. Press Accept.