PAGE 16 OF 54
0521 IH-5276
OPERATION CONTINUED
TOTALIZATION
Totalization measures the cumulative weight of a
sequence of items. (See Figure 62) The cumulative
total may exceed the capacity of the balance. The
maximum number of samples (n) is 99.
1. In the upper left portion of the home screen, select
Totalization.
2. Place item on the pan to begin. The sample weight
is shown on the main display line.
3. Press
ACCUMULATE
to add the weight (when stable)
of the item to the total.
4. Remove the item from the weighing pan, then add
the next item and continue as above.
5. Press
RESULT
to view the results from the totalization.
(See Figure 63)
6. When finished, press the Clear Total button to reset
the accumulated weight to zero.
7. Press
SAVE TO USB
to save the result to a USB flash
drive or
EXIT
to return to the totalization home screen.
NOTE: Changing units converts the
accumulation results to the selected unit.
FORMULATION
Use this application for compounding and recipe
making. (See Figure 64) The maximum number of
components is 50.
1. In the upper left portion of the home screen, select
Formulation. (See Figure 65)
2. Press
START
to begin the compounding process.
3. Place the first ingredient on the pan (or in a
tared container) and press
ACCEPT
to store the
component.
4. Continue adding components and pressing
ACCEPT
to store the weight of the individual components
until the formula is complete. The
Total
line shows the
total weight of all the components.
5. Press
STOP
to finish the formulation. The formulation
results are displayed:
NOTE: If Filler is set to On (see Item Settings on
page 17), the balance will ask to add a filler
material to complete the formulation. Add the
filler material and press ACCEPT to complete
the formulation and display the results.
6. Press
SAVE TO USB
to save the results to a USB flash
drive or
EXIT
to return to the Formulation Home screen.
NOTE: The formulation results will be cleared
when a new formulation starts.
Totalization
Exit
Item
Samples
Total
Range
Average
Minimum
Maximum
Save to USB
o (stdev)
Result
3
506.95 g
93.57 g
168.98 g
117.00 g
210.57 g
38.90 g
The
TOTALIZATION
home screen
Main Display Line
Function Buttons
Reference Fields
70.19
Totalization
Menu
Clear Total
Accumulate
Result
70.19 g
0.00 g
Gross:
Tare:
0
0.00 g
Samples:
Total:
>0<
g
Max 4200 g
d=0.01 g
Place sample on the pan. Press Accumulate to add to the total.
Figure 62
Figure 64
Figure 63
Figure 65
The
FORMULATION
home screen
Main Display Line
Function Buttons
Reference Fields
0.00
Formulation
Menu
Item
Settings
Start
0.00 g
0.00 g
Gross:
Tare:
Off
On
Filler:
Auto Tare:
>0<
g
Max 4200 g
d=0.01 g
Press Start to begin Formulation.
Formulation
Exit
Comp.
1
2
Comp. Total
Save to USB
Weight
241.76 g
272.95 g
514.71 g