PAGE 7 OF 54
0521 IH-5276
ITEM SETTINGS CONTINUED
The next screen appears, with the message to place the
reference weight on the pan. (See Figure 17)
Place the reference weight on the pan, then touch
ACCEPT
to capture the value; the screen shows number
of pcs. (See Figure 18)
Establish an Average Piece Weight (APW):
Each time a
new type of part is counted, the nominal weight of one
piece (APW) must be established using a small quantity
of pieces. This APW is stored until replaced by another
APW.
There are two methods to establish the APW value:
1. The actual APW is known.
2. The APW must be derived by weight. For this case
the current sample size will be used. (See Figure 19)
Set a Known Average Piece Weight (APW):
To adjust the
APW value directly, touch the
APW
button. A numeric
input window appears. (See Figure 20)
Key in the desired APW Weight, then press
SAVE
.
The display returns to the application home screen with
the new APW value in the reference field. (See Figure 19)
Set a New Average Piece Weight (APW) – Derived:
To
establish a new APW, place the reference weight on the
pan and press
ACCEPT
to continue. (See Figure 21)
NOTE: The sample size that is displayed will be
used. To use a different sample size, change
that first. (See Figure 21)
The home screen shows the number of pieces at the
new APW. (See Figure 22)
OPERATION CONTINUED
Figure 18
Parts Counting
Exit
Back
Accept
APW
0.010 g
0.00 g
0.00 g
Gross:
Tare:
0.010 g
10
APW:
Sample Size:
Place reference weight on the pan. Press Accept to continue.
>0<
0.00
g
Max 4200 g
d=0.01 g
Parts Counting
Menu
Item
Settings
Sample
Name
Save to
USB
70.19 g
0.00 g
Gross:
Tare:
3.510 g
Off
APW:
Auto Optimization:
20
PCS
Max 4200 g
d=0.01 g
Figure 19
Parts Counting
0.00 g
0.00 g
Gross:
Tare:
0.010 g
10
APW:
Sample Size:
Place reference weight on the pan. Press Accept to continue.
>0<
0.00
g
Max 4200 g
d=0.01 g
Exit
Back
Accept
APW
0.010 g
APW
0.010 g
1
2
3
4
5
6
7
8
9
Clear
0
•
Exit
Save
Figure 20
Figure 21
Parts Counting
Exit
Back
Accept
APW
0.010 g
0.00 g
0.00 g
Gross:
Tare:
0.010 g
10
APW:
Sample Size:
>0<
Place reference weight on the pan. Press Accept to continue.
0.00
g
Max 4200 g
d=0.01 g
Figure 22
Parts Counting
Menu
Item
Settings
Sample
Name
Save to
USB
179.75 g
0.00 g
Gross:
Tare:
8.987 g
Off
APW:
Auto Optimization:
20
PCS
Max 4200 g
d=0.01 g
Figure 17