0415 IH-2709

PAGE 4 OF 8

ENTERING THE .U.N.I.T. MENU 

The .U.N.I.T. menu contains units of measure, PC (parts 

counting), % weighing and END. Units vary with the model 

type. Determine which units are to be turned on or off. 

Start in the .U.N.I.T. menu. Select either 

ON or OFF for 

each unit. 

NOTE: Repeated presses of PRINT Unit button 
will go through all units, you then may select 

ON or OFF. Parts Counting is slightly different.

PARTS COUNTING 

Two types of counting modes are available: standard or 

enhanced. 

EXITING THE .U.N.I.T. MENU 

Use the same procedure as exiting the .S.E.T.U.P. Menu.

OPERATION CONTINUED

1

1

1

1

1

1

Y

Y

N

N

N

N

CALIBRATION 

ULINE MODEL NO.

H-2709

H-2710

H-2111

H-2382

H-2383

H-2384

H-2712

H-3054

H-2713

OHAUS MODEL NO.

SPx202

SPx402

SPx602

SPx401

SPx601

SPx2001

SPx4001

SPx6001

SPx6000

Span Calibration Mass (g)

0, 200

0, 200

0, 300

0, 200

0, 300

0, 2000

0, 4000

0, 6000

0, 4000

Linearity Calibration Mass (g)

0, 100, 

200

0, 200, 

400

0, 300, 

600

0, 200, 

400

0, 300, 

600

0, 1000, 

2000

0, 2000, 

4000

0, 3000, 

6000

0, 3000, 

6000

CALIBRATION VALUES

1

1

1

1

1

Y

Y

Y

N

N

(Turn on)

(To accept)

(To advance)

SPAN CALIBRATION

Span calibration uses two calibration points, zero and a 

specified calibration weight. 

Before beginning calibration, make sure the Lock Switch 

is unlocked. Clear the platform. 

NOTE: Value of calibration mass depends on 

capacity of balance. After calibration, the balance 

returns to the currently selected weigh mode. 

1.  Start with the balance OFF. 

2.  Hold the 

ON/ZERO Off button until the display reads 

MENU CAL

3.  Press and release 

ON/ZERO Off. Scale will flash 

-C-

 and show required weight. Place the calibration 

weight on the scale.

4.  Press and release 

ON/ZERO Off. Scale will show 

donE

; remove weight.

•••

5

1

Y

(Acquiring zero value)

(Place mass value

on platform)

1

Y

(Remove mass)

Summary of Contents for SCOUT PRO H-2709

Page 1: ...r the button is held for 3 seconds In Display Hold or Totalize mode a long press exits the more without turning the balance off UNIT Press and hold scrolls through units Release on desired unit MENU F...

Page 2: ...BATTERY INSTALLATION Install 4 AA batteries in the battery compartment See Figure 5 AC ADAPTER INSTALLATION Plug the AC adapter into the jack at the rear of balance See Figure 5 Figure 5 Battery Comp...

Page 3: ...entering the Mode submenu Starting from the C A L menu NOTE Select ON or OFF for Auto Off To enter Display Hold or Totalize you must first enter the Mode sub menu TURNING DISPLAY HOLD OR TOTALIZE MODE...

Page 4: ...Px6001 SPx6000 Span Calibration Mass g 0 200 0 200 0 300 0 200 0 300 0 2000 0 4000 0 6000 0 4000 Linearity Calibration Mass g 0 100 200 0 200 400 0 300 600 0 200 400 0 300 600 0 1000 2000 0 2000 4000...

Page 5: ...t remains until ON ZERO Off button is pressed again or the balance is turned off PARTS COUNTING Parts Counting is enabled only when PC is turned ON in the U N I T menu In parts counting mode there are...

Page 6: ...ht equals 100 Start in the weighing mode and zero the display Place load on the platform to display percent of reference weight ESTABLISHING A NEW REFERENCE WEIGHT Press and hold until on the display...

Page 7: ...ally elevated supported on all feet and leveled A fine wire is attached to the built in hook at the bottom of the balance LOCK SWITCH The Lock Switch is located under the balance The Lock Switch is us...

Page 8: ...ces nearby affecting operation Either move or shield the balance from external air currents or vibration Err 1 chEct Invalid checksum data Cycle balance ON OFF Return balance for service Err 2 LoAd Ov...

Reviews: