0415 IH-2709
PAGE 4 OF 8
ENTERING THE .U.N.I.T. MENU
The .U.N.I.T. menu contains units of measure, PC (parts
counting), % weighing and END. Units vary with the model
type. Determine which units are to be turned on or off.
Start in the .U.N.I.T. menu. Select either
ON or OFF for
each unit.
NOTE: Repeated presses of PRINT Unit button
will go through all units, you then may select
ON or OFF. Parts Counting is slightly different.
PARTS COUNTING
Two types of counting modes are available: standard or
enhanced.
EXITING THE .U.N.I.T. MENU
Use the same procedure as exiting the .S.E.T.U.P. Menu.
OPERATION CONTINUED
1
1
1
1
1
1
Y
Y
N
N
N
N
CALIBRATION
ULINE MODEL NO.
H-2709
H-2710
H-2111
H-2382
H-2383
H-2384
H-2712
H-3054
H-2713
OHAUS MODEL NO.
SPx202
SPx402
SPx602
SPx401
SPx601
SPx2001
SPx4001
SPx6001
SPx6000
Span Calibration Mass (g)
0, 200
0, 200
0, 300
0, 200
0, 300
0, 2000
0, 4000
0, 6000
0, 4000
Linearity Calibration Mass (g)
0, 100,
200
0, 200,
400
0, 300,
600
0, 200,
400
0, 300,
600
0, 1000,
2000
0, 2000,
4000
0, 3000,
6000
0, 3000,
6000
CALIBRATION VALUES
1
1
1
1
1
Y
Y
Y
N
N
(Turn on)
(To accept)
(To advance)
SPAN CALIBRATION
Span calibration uses two calibration points, zero and a
specified calibration weight.
Before beginning calibration, make sure the Lock Switch
is unlocked. Clear the platform.
NOTE: Value of calibration mass depends on
capacity of balance. After calibration, the balance
returns to the currently selected weigh mode.
1. Start with the balance OFF.
2. Hold the
ON/ZERO Off button until the display reads
MENU CAL
3. Press and release
ON/ZERO Off. Scale will flash
-C-
and show required weight. Place the calibration
weight on the scale.
4. Press and release
ON/ZERO Off. Scale will show
donE
; remove weight.
•••
5
1
Y
(Acquiring zero value)
(Place mass value
on platform)
1
Y
(Remove mass)