0415 IH-2709
PAGE 6 OF 8
ENHANCED PARTS COUNTING
In the enhanced mode, additional parts can be added
to the platform equal to or less than the original number.
The additional reference quantity produces a more
accurate average part weight.
NOTE: Enhanced Parts Counting can be
repeated as many times as necessary
providing the quantity added is less than the
original entry.
To count different parts, press and hold, until PC
reappears, release
Unit button.
PERCENT WEIGHING
Percent Weighing is enabled only when Percent is
turned ON in the .U.N.I.T. menu. Percent weighing
permits placing a reference weight on the balance,
then viewing other loads as a percentage of the
reference. The reference weight equals 100%.
Start in the weighing mode and zero the display.
Place load on the platform to display percent of
reference weight.
ESTABLISHING A NEW REFERENCE WEIGHT
Press and hold until % on the display reappears, then
release.
Repeat above procedure for new a reference weight.
EXITING PERCENT WEIGHING
Press and hold until desired unit is displayed.
DISPLAY-HOLD
Display-Hold is enabled only when Hold is turned ON in
the Mode submenu in the .S.E.T.U.P. menu. Display-Hold
mode captures and stores the highest stable value.
When displayed, the stable icon will blink.
NOTE: Units cannot be changed when in
display hold mode.
EXITING DISPLAY HOLD
To return to display-hold, repeat above procedure.
OPERATION CONTINUED
(Add less than 50)
(Add parts to be counted)
1
1
1
Y
Y
Y
(Press and hold until % is displayed)
(Place reference
weight)
(Stores reference
weight)
(Remove reference weight)
1
(Place item(s) on platform)
3