background image

0415 IH-2709

PAGE 8 OF 8

OPERATING ISSUE

CAUSES

RECOMMENDATIONS

No display.

Power Adapter not connected.

Batteries are exhausted.

Connect AC adapter.

Replace batteries.

Battery Indicator is flashing.

Batteries are weak.

Replace batteries.

Incorrect weight reading.

Balance out of calibration.

Balance was not rezeroed before 

weighing.

Calibrate the balance. 

Press the 

ON/ZERO Off button with no weight 

on the platform, then weigh item. 

Calibration procedures do 

not work.

Incorrect calibration masses being used.

Use correct masses.

See error codes note ERR4.

Unable to switch unit of 

measure.

Weighing unit not activated in menu. 

Mode prevents unit change.

Use Units menu to set desired units ON.

Exit Hold or Totalize mode.

Balance readings unstable.

Balance location may have drafts from 

air conditioning vents or vibration from 

other appliances nearby affecting 

operation.

Either move or shield the balance from 

external air currents or vibration.

Err 1 chEct 

Invalid checksum data.

Cycle balance ON/OFF.

Return balance for service.

Err 2 LoAd

Overload (>cap+90d) or Underload.

Remove load.

Make sure shipping lock is unlocked.

Recalibrate.

Err 4 WEIght 

Incorrect calibration weight.

Use correct calibration weights.

Check environmental conditions.

Err 5 999999 

Displayed value>99999.

Exit application.

Rezero balance.

Err 6 count 

Parts counting error – balance exits  

parts counting.

Exit application.

Rezero balance.

Err 8 FULL 

RS232 buffer is full.

Cycle ON/OFF.

Check RS-232 settings in balance and 

computer.

Err 9 dAtA 

Internal data error. The next button 

press or a 5 second time-out causes 

the balance to turn off.

Cycle ON/OFF.

Return balance for service.

TROUBLESHOOTING

π

  CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA

1-800-295-5510

uline.com

 If the troubleshooting section does not resolve your problem, contact ULINE Customer Service at 1-800-295-5510.

Summary of Contents for SCOUT PRO H-2709

Page 1: ...r the button is held for 3 seconds In Display Hold or Totalize mode a long press exits the more without turning the balance off UNIT Press and hold scrolls through units Release on desired unit MENU F...

Page 2: ...BATTERY INSTALLATION Install 4 AA batteries in the battery compartment See Figure 5 AC ADAPTER INSTALLATION Plug the AC adapter into the jack at the rear of balance See Figure 5 Figure 5 Battery Comp...

Page 3: ...entering the Mode submenu Starting from the C A L menu NOTE Select ON or OFF for Auto Off To enter Display Hold or Totalize you must first enter the Mode sub menu TURNING DISPLAY HOLD OR TOTALIZE MODE...

Page 4: ...Px6001 SPx6000 Span Calibration Mass g 0 200 0 200 0 300 0 200 0 300 0 2000 0 4000 0 6000 0 4000 Linearity Calibration Mass g 0 100 200 0 200 400 0 300 600 0 200 400 0 300 600 0 1000 2000 0 2000 4000...

Page 5: ...t remains until ON ZERO Off button is pressed again or the balance is turned off PARTS COUNTING Parts Counting is enabled only when PC is turned ON in the U N I T menu In parts counting mode there are...

Page 6: ...ht equals 100 Start in the weighing mode and zero the display Place load on the platform to display percent of reference weight ESTABLISHING A NEW REFERENCE WEIGHT Press and hold until on the display...

Page 7: ...ally elevated supported on all feet and leveled A fine wire is attached to the built in hook at the bottom of the balance LOCK SWITCH The Lock Switch is located under the balance The Lock Switch is us...

Page 8: ...ces nearby affecting operation Either move or shield the balance from external air currents or vibration Err 1 chEct Invalid checksum data Cycle balance ON OFF Return balance for service Err 2 LoAd Ov...

Reviews: