background image

14

5.2 Unit Switching

The USS-DBS92 Bench Scale offers two weight units of

kg

and

lb

. The

unit selection function can be used to meet the unit requirements in

various usage situations.

In the weighing state, press the "UNIT" button to switch weight units,

the

weighing unit indicator - lb

and

weighing unit indicator - kg

indicates current unit.

5.3 Counting

The U.S. Solid USS-DBS92 Bench Scale offers a counting function that

meets various industrial counting requirements, especially the counting

function for small components.

To ensure the accuracy of the counting function for small components, it

is necessary to ensure that the weight of all samples is consistent, and

that the weight of a single object is

40g.

1. Press

COU

and

-COU-

will be displayed;

2. After that a quantity will be flashing displayed;

3. Press

TARE

to select the quantity and put the corresponding

Summary of Contents for USS-DBS92

Page 1: ...Instruction Manual USS DBS92 Bench Scale ...

Page 2: ... 3 1 Unpacking and delivery inspection 8 3 2 Installation 9 Before Using 10 4 1 Leveling the scale 10 4 2 Calibration 12 Operation 13 5 1 Basic Weighing 13 5 2 Unit Switching 14 5 3 Counting 14 Setting 16 6 1 Minimum Weighing 16 6 2 Baud Rate 16 Maintenance 17 7 1 Precautions 17 7 2 Cleaning 18 7 3 Disposal 18 Troubleshooting 19 Technical Data 20 ...

Page 3: ...manual completely and follow the usage instructions before installation and operation as this will help you to make full use of the functions and performance of the USS DBS92 Bench Scale 1 1 Safety Precautions The U S Solid USS DBS92 Bench Scale qualifies as state of the art technology and complies with all recognized safety rules Improper use or handling however can result in damage and or injury...

Page 4: ...f excessive voltage is used the scale may overheat and be damaged or cause a fire Operate the scale on a stable rigid and flat table Handle the scale carefully It is a precision device subjecting it to impact may result in a malfunction If the device is not be used for a long period of time the power should be turned off and disconnect the power cable Do not disassemble remodel or repair this prod...

Page 5: ...3 Design and Function 2 1 Components 1 Weighing Indicator 4 Scale Platform 2 Mounting Bracket 5 Scale Base 3 Column Assembly 6 Leveling Feet ...

Page 6: ...4 7 RS232 Interface 8 Power Switch 9 Power Inlet 10 Connecting Line Interface 11 Rechargeable Battery Cover ...

Page 7: ...tting of baud rate Perform weight calibration Tare or zero the scale Perform the baud rate set while switching on Switch the option in counting mode Enter the counting function Perform the minimum weighing setting Select the option in the setting of baud rate ...

Page 8: ...ltage indicator Indicates the remaining battery power is insufficient and needs to be charged 3 Stability indicator Indicates stabilization of scale 4 Counting indicator Indicates current working mode is counting mode 5 Weighing unit lb s Indicates current unit is lb S 6 Weighing unit kg Indicates current unit is kg ...

Page 9: ...7 7 Power indicator Charging red light Fully charged critical point red and green lights alternately Fully charged green light No battery but power connected green light ...

Page 10: ...ssories are part of the standard equipment for your new USS DBS92 Bench Scale 1 Scale Platform and Base 1 Weighing Indicator 1 Column Assembly 1 Mounting Bracket 4 Leveling Feet 1 AC Adapter 1 Rechargeable Battery 1 Instruction Manual 1 Quality Certification Check the instrument for damage in transit Immediately inform the U S Solid customer service if you have any complaints or parts are missing ...

Page 11: ...nd leveling feet into the screw holes on the four corners at the bottom of the scale base Connect the mounting bracket to the column and then tighten the screw Pass the connecting line of the scale base through the column and the mounting bracket to connect it to the back of the weighing indicator Insert the mounting bracket under the weighing indicator Put the column vertically into the connectin...

Page 12: ...ery capacity is sufficient and the power cord needs to be connected 4 1 Leveling the scale Accurate horizontal positioning is necessary for repeatable measurements and exact results To compensate for small irregularities or tilts at this location the instrument needs to be leveled Adjust the leveling feet of the bench scale until the air bubble in the indicator is centered The level indicator is l...

Page 13: ...of air bubble Adjustment method up Turn both feet clockwise down Turn both feet counterclockwise Left Turn left foot counterclockwise right foot clockwise Right Turn left foot clockwise right foot counterclockwise ...

Page 14: ... Changes in ambient conditions 1 Press TARE to zero the scale in the no load state 2 Long press the CAL button and CAL will be displayed 3 After that 200 00 or 50 00 will be flashing displayed then place the calibration weight on the pan 4 will be displayed then 200 00 or 50 00 will be displayed again 5 Remove the calibration weight and will again displayed 6 The calibration is completed and it wi...

Page 15: ...ing container on the weighing pan 2 Wait for the display to stabilize Then press TARE to tare and the displayed value will return to zero 3 Place the sample s to be weighed in the weighing container 4 After the display stabilizes read the value displayed Overweight Alarm When is displayed in the weighing state it indicates that the sample weight exceeds the maximum range of the scale The sample sh...

Page 16: ...ent unit 5 3 Counting The U S Solid USS DBS92 Bench Scale offers a counting function that meets various industrial counting requirements especially the counting function for small components To ensure the accuracy of the counting function for small components it is necessary to ensure that the weight of all samples is consistent and that the weight of a single object is 40g 1 Press COU and COU wil...

Page 17: ...ss COU to confirm and will be displayed 5 Put samples to be counted and the value of corresponding quantity appears Exit counting function 1 Press COU and will be displayed 2 Then it will exit the counting function and return to the weighing interface ...

Page 18: ...turn up or turn down respectively from 1d to 5d d means minimum scale value 10 g 3 Press TARE to confirm 6 2 Baud Rate The baud rate could be set to change the transmission speed of the interface 1 Press TARE while switching on and the baud rate options will be displayed cyclically 2 When looping to the required baud rate press COU to confirm the selection 3 Then the baud rate set successfully ...

Page 19: ... to the weighing system do not let the object fall from the height onto the weighing pan 3 Do not expose the scale to high humidity or dusty conditions for an extended period of time 4 When the scale is not to be used for a long time cover it with to prevent dust from entering 5 When weighing powder and fine particles should be done with a suitable container to prevent dust and particles from fall...

Page 20: ...eaning agents solvents abrasive cleaning agents etc or organic solvents to clean the scale Clean the scale using a piece of lint free cloth which has been wet with a mild detergent soap only 7 3 Disposal Disposal of equipment and packaging must be carried out by the operator in accordance with relevant laws of the country or region in which the equipment is to be used ...

Page 21: ...ower transformer damaged Instrument faulty Problem Weight display is constantly changing or unstable Possible causes Test bench vibration Electromagnetic field interference Power supply instability Problem Displayed value and actual weight are not consistent Possible causes Not returned to zero before weighing Scale not leveled well ...

Page 22: ...ity 160 kg 310 kg Readability 10g Repeatability 20 g Linearity Deviation 30 g Stabilization Time Within 3 seconds Pan Size 16 20 inches 400 500 mm Interface RS232 Power 110V 60Hz AC Dimensions 24 4 15 8 33 9 inches Net Weight 30 9 lbs Shipping Weight 35 3 lbs ...

Page 23: ......

Page 24: ...Contact Feel free to visit our website www ussolid com You can email us at service ussolid com You can call one of our friendly customer service representative at 1 800 209 4177 ...

Reviews: