X.
Warnings and Notes
1.
NEVER START ULTRASONIC OUTPUT WHEN THE PROBE TIP
IS
EXPOSED IN
THE AIR
2.
The apparatus
swithes the
power supply without industrial frequency
transformer. Don’t randomly
touch
after opening the generator housing to
avoid electr
ic
shock.
3.
The apparatus does not need
to
warm up and should be grounded well
when
being used.
4.
Use it in
an
environment free from moisture, sunshine and corrosive
gases.
5.
Be sure the platform is tightly fixed to the rod,
and
the ultrasonic probe is
in
liquid and not touching the walls of the sample vessel.
6.
The liquid temperature will increase rapidly due to the cavitation effect
during crushing. The user should pay attention to the temperatures for
different cells. It is suggested
that the user
adopt multiple short time
crushing (not
longer than 5s each time) and ice bath cooling.
7.
It is
practice-proven
that multiple short
-
ime operations, working
time 1
-
2s
and interval time 1
-
2s, have better effects than continuous long
-
time
operations. The long interval time can be set to avoid the heating of
liquid.
Additionally, continuous long
-
time operations are subject to no load
operations to shorten the apparatus service lifetime.
8.
The horn end will become rough due to cavitation corrosion after being
used for a period of time. Use oil stone or rasper to smooth
it
. Otherwise,
it
will influence the working effects.
9.
In general, the volume of homogenization media should be 3 to 10 times
the net volume of the solid sample. Pre-chopping the sample into pieces
less than 1 mm in cross-section will significantly decrease the time of
homogenization.