VII. Energy Converter Transducer
3. Piezoelectric Ceramics
1. High Tensile Screw
4. Amplitude
2. Electrode Plate
5. Connecting Screw
VIII. Installation
1.
Have the bracket and clamp assembled (the holding system for
the
transducer).
2.
Have the transducer cable and transducer connected, then connect
them
with
the main machine (ultrasound generator).
3.
Connect
the
power cable to
the
power supply and main machine.
4.
Put the transducer unit on the bracket or
soundproof box
. Make
sure the
ultrasonic probe tip is inserted into the sample liquid surface and
in the
center of
the
container.
The probe
should get into the liquid 5
mm to
10
mm
beneath the liquid surface.
For large sample volumes, the distance
between the probe tip and the container’s bottom should be bigger than
30
mm. (The ultrasonic probe tip can be as close as 10
mm from the
bottom
of the container when processing a small sample
size
or when
using a
lower transducer power setting) If you hold the transducer unit to
process
the sample, hold it on SLEEVE position. Other positions are not
allowed.
It is prohibited to start ultrasonic output when the probe tip is
out of the liquid
. O
therwise the energy converter and
ultrasonic wave generator will be damaged.