37
5.
Note the level (H or L) in the TG box of the Sensitivity Validation chart.
6.
With the handle facing the user, align the tip of the contact
probe with the recharge jack and allow the probe to rest on
the jack. DO NOT PRESS DOWN!
7.
(NOTE: NEVER USE THE ALUMINUM EXTENSION PROBE
RODS THEY WILL SHORT OUT THE BATTERY OF THE WTG).
8.
Adjust the sensitivity until intensity bar graph is at mid-line.
9.
Note and record the decibel in the dB box of the Sensitivity
Validation chart.
For all tests
:
When performing a Sensitivity Validation Test, review the data in the Sensitivity
Validation chart and repeat the test using the same rod/module, frequency, and
Warble Tone Generator volume setting.
Look for a change in the decibel reading. A change of greater than 6 dB will indicate a
problem.