SAFETY INSTRUCTIONS

INTENDED USE

•  Please read through these instructions carefully before using the thermometer.

•  Measuring your temperature for self-diagnosis is dangerous! Please consult a doctor 

for an interpretation of the measurement readings. Self-diagnosis can aggravate an 

existing medical condition.

•  This device is not suitable for temperature measurement in the ear.

This digital thermometer is intended to measure the human body temperature in regular 

mode orally (in the mouth), rectally (in the anus) or axillary (under the arm). It is reusable 

for clinical or home use on people of all ages.

Digital Medical Thermometer

PZN 10407062

Instructions for use

0123

0816

Tip

Display

On/off button

Battery compartment cover

•  For rectal measurement, do not insert the measuring tip with force. Discontinue the 

measurement if you feel any pain. Risk of injury!

•  Do not use the thermometer for oral measurements if it has previously been used for 

rectal measurements.

•  Do not conduct oral measurements on children under the age of 2.

•  Do not bite the thermometer. This can cause damage or injury!

•  Do not drop the thermometer. It is not resistant to shock or impact.

•  Where protective coverings are used, the temperature measurement can deviate by 

around 0.1°C.

•  The thermometer contains small parts (battery etc.) which children could swallow. Do 

not let children use the device unsupervised.

•  Protect the device from high temperatures and direct sunlight.

•  Do not expose the battery to extreme heat. Risk of explosion!

•  Use outside the intended operating conditions or storage outside the intended storage 

and transport conditions can result in inaccurate measurements.

•  Store the thermometer in the transparent box when not in use.

•  Remove the battery when the device is not being used for a long period.

•  Using the unit in the immediate vicinity of mobile phones, microwave appliances or 

other devices with strong electromagnetic fields may result in impaired functioning. 

Maintain a minimum distance of 3 m from such devices when using this unit.

•  Do not open the casing (except to change the battery).

Warning: Always consult a doctor if the reading shows a high fever.

MEASURING THE TEMPERATURE

The skin must be completely dry. Position the measuring tip under your arm in such a 

way that the measuring tip is in good contact with your skin. Press your arm closely to 

your  body.  From  a  medical  perspective,  the  axillary  measurement  method  produces 

inaccurate measurement readings and should not be used if you wish to take precise 

measurements.

In the rectum (rectal measurement)

Apply some Vaseline to the measuring tip to aid insertion. The measuring tip is carefully 

inserted 1 cm into the anus.

Insert the tip into one of the two heat pockets under the tongue, on the left- or right-hand 

side of the root of the tongue. Close the mouth and breathe calmly through the nose.

Under the arm (axillary measurement)

In the mouth (oral measurement)

Place  the  thermometer  at  the  desired  measurement  point.  During  measurement  the 

current temperature is shown on the display with a flashing “°C” symbol. A beep sound 

signals the end of the measurement. It is imperative you adhere to the minimum meas-

urement time denoted by this signal tone. Please note that the thermometer will continue 

to measure after the signal tone.

To turn on the thermometer, press the on/off button. A short beep indicates that the 

thermometer is on. At the same time, an optical full segment check of 

the display will be carried out. Then the last temperature measured will 

be displayed for approx. 2 seconds, along with a small M (for memory). 

Then  an  internal  test  value  appears  and  the  thermometer  switches  to 

measuring  mode.

Read the temperature from the display. The thermometer will switch itself off automat-

ically after approx. 10 minutes. You can also turn it off earlier by pressing the on/off 

button.

CLASSIFICATION OF THE TEMPERATURES MEASURED

MEASUREMENT TYPES

For temperatures up to 37.8 °C, the normal sound will be made (10 long, equal beeps). 

For temperatures over 37.8 °C, the fever alarm will sound (30 short beeps in groups of 3).

Body temperature measured orally is on average between 35.7 and 37.3 °C. For temper-

ature classification, please refer to the following approximate values:

37.3 °C to 37.7 °C: elevated temperature 

37.8 °C to 38.9 °C: moderate fever 

39.0 °C and above: high fever

Temperatures measured rectally are generally 0.5 °C higher than those measured orally, 

while temperatures measured under the arm are 0.5 °C lower.

CHANGING THE BATTERY

When the display shows the   symbol, the battery needs to be changed.

CLEANING

DISPOSAL

METROLOGICAL INSPECTION

•  Insert a new battery (LR41 or SR41, 1.5V) with the + sign towards the 

top.

Clean the thermometer before and after every use, using a soft cloth and isopropyl alco-

hol diluted with water, or cold soapy water.

Batteries and technical appliances must not be disposed of with domestic waste, 

but should be handed in at the appropriate collection and disposal points.

For cleaning, do not use petrol, thinner or any other strong solvent.

Do not immerse the device in alcohol for a prolonged period. Do not expose the ther-

mometer to temperatures over 50 °C. Do not use ultrasonic cleaning.

Basically, a metrological inspection is recommended at intervals of 2 years. However, 

professional users in Germany have to comply with the aforementioned according to 

“Regulation for Operators of Medical Devices”. This can be implemented either by UEBE 

Medical GmbH, an authority responsible for metrology, or authorised maintenance ser-

vices. For this, please observe your national provisions.

•  Push the battery holder back into the casing and replace the battery 

compartment cover. Take care not to damage or misalign the seal when 

doing this.

EXPLANATION OF SYMBOLS

This product complies with the Council Directive 93/42/EC from 5 Septem-

ber 2007 regarding medical devices and bears the mark CE 0123 (TÜV SÜD 

Product  Service  GmbH).

Storage and transportation conditions  

Ambient temperature -20 to 55 °C, rel. air humidity 15 to 95 %

Degree of protection against electric shock: TYP BF

Please observe operating instructions

Lot number

Manufacturer

DC direct current

55 °C

-20 °C

95 %

15 %

•  Carefully pull the battery holder about 1 cm out of the casing.

•  Please use a non-metallic pointed object to remove the battery from 

its  holder.

•  Remove the battery compartment cover from the thermometer.

TECHNICAL DETAILS

ERROR MESSAGES

Measured temperature is below 32.0 °C and thus outside the measure-

ment range.

Measured temperature is above 42.9 °C and thus outside the measure-

ment range.

Electronic error. If error persists, please contact UEBE customer service.

Type:

Maximum thermometer, direct mode

Range:

32.0 °C to 42.9 °C

Accuracy:

± 0.1 °C between 35.5 °C and 42.0 °C 

± 0.2 °C in other temperature ranges

Operating conditions: Ambient temperature 15 to 35 °C, relative humidity 15 to 95 %
Storage and transport 

conditions:

Ambient temperature -20 to 55 °C, relative humidity 15 to 95 %

Battery:

LR41 or SR41, 1,5 V

Battery life:

Approx. 200 hours

IP rating:

IP 27: protection from solid foreign bodies with diameters of 

12.5mm and above, protection against temporary submersion

WARRANTY

During the warranty period of 2 years from the date of purchase we reserve the right 

either to repair any such defect at our expense or to supply a perfect replacement unit. 

The cost of returning the unit to our factory shall be borne by the sender. UEBE shall 

refuse to accept return deliveries that have not been paid for by the sender.

Excluded from the warranty are parts subject to normal wear and tear as well as dam-

age caused by non-compliance with the instructions for use, improper handling (e.g. 

unsuitable power sources, breakages, leaking batteries) and/or disassembly of the unit 

by the purchaser. Furthermore, no claims for damages against us are substantiated by 

the warranty.

Warranty claims can only be advanced in the warranty period and by presenting proof of 

purchase. In the event of a warranty claim, the unit must be sent to the following address 

together with the proof of purchase and a description of the complaint: Công ty TNHH 

YSMART, P1104-Nhà C5-Khu Đô Thị Mỹ Đình 1 - Q.Nam Từ Liêm - Hà Nội – Việt Nam.

The statutory claims and rights of the buyer against the seller (claims for defect, manu-

facturer’s liability, for example) are not restricted by this warranty.

Please note: In the event of a warranty claim it is essential to attach the proof 

of purchase.

© 2014 UEBE Medical GmbH, Not to be re-

produced, either in whole or in part, subject to 

technical  modifications

7 0816 207 AA  2017-07

UEBE Medical GmbH 

Zum Ottersberg 9 

97877 Wertheim, Germany

Công ty TNHH YSMART 

P1104-Nhà C5-Khu Đô Thị Mỹ Đình 1 - 

Q.Nam Từ Liêm - Hà Nội – Việt Nam 

Tel: +84. 04 62871707 

Email: info@uebe.vn 

Internet: http://uebe.vn

Reviews: