background image

13

REPLACING THE BATTERIES

POST TEST

PERIODIC SERVICE

ANNUAL RE-CERTIFICATION

CLEANING

This meter has been designed for use with both alkaline and rechargeable Nickel Metal Hydride (NiMH) batteries . No other types are 

recommended . The analyzer is supplied with 4 “AA” size alkaline batteries . These should be installed into the instrument as shown in the diagram 

indicated on the back of the unit .

 CAUTION!

 

Take great care when installing the batteries to observe correct polarity .  

 

 

 

 

 

 

Always check the meter for operation immediately after installing    

 

 

 

 

 

 

new batteries .

Using Re-Chargeable Batteries

The battery charger must only be used when NiMH batteries are fitted .  Alkaline batteries are not re-chargeable . Attempting to recharge alkaline 

batteries may result in damage to the product and may create a fire risk .

Battery Charging

Ensure that you use the correct charger . This unit uses a 9V DC regulated charger . Ensure that the batteries are fitted in the correct manner, and 

then charge for at least 16 hours . Subsequent charges should be overnight . NiMH batteries may be charged at any time, even for short periods 

to conduct testing .

 

WARNING!

 

Under NO circumstance should you expose batteries to extreme heat or fire as they may explode and cause injury . Always  

 

dispose of old batteries promptly in a manner consistent with local disposal regulations .

WARNING!

Repair and service of this instrument is to be performed by qualified personnel only . Improper repair or service could result in 

physical degradation of the instrument . This could alter the protection from personal injury this meter provides to the operator . 

Perform only those maintenance tasks that you are qualified to do .

Remove the probe from the flue and allow the analyzer to purge 

with fresh air until readings return to zero . - O2 to 20 .9%, CO to 

Zero (Be careful as the probe tip will be HOT)

C85

L O N G   L I F E   E O S   A N A L Y Z E R

CO

2

 CO TEMP

Menu

Status

Aux

Eff 

X-Air

O2

CO Air

Free

       CO

2

CO

Temp

Fuel

Date

Time

O2% 20.9

COa -0>-

Drain water trap by unplugging 

the drain plug and shake to 

get excess water out .

Check particle filter for dirt 

and any other sediment and 

replace if necessary .

While the CO sensor has an expected life of more than five years in normal use it is recommended that the analyzer is re-certified at least annually . 

This is so that long-term drift on the sensor and electronics can be eliminated . Local regulations may require more frequent re-calibration and 

users should check with appropriate authorities to ensure they comply with relevant guidelines .

Periodically clean your instruments case using a damp cloth . DO NOT use abrasive, flammable liquids, cleaning solvents, or strong detergents as 

they may damage the finish, impair safety, or effect the reliability of the structural components .

www.

.com

information@itm.com

1.800.561.8187

Summary of Contents for 053533507225

Page 1: ...I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time COp 28 CO2 6 2 C85 INSTRUCTION MANUAL ENGLISH Long Life EOS Analyzer www com information itm com 1 800 561 8187 ...

Page 2: ...odic Service 13 Analyzer Re Certification 13 Cleaning 13 Other Important Factors Related To Combustion 14 Combustion Measurement Terms 14 Specifications 15 Electromagnetic Compatibility 15 Limited Warranty 16 DISPLAY SYMBOLS Excess Air Theoretical amount of air in excess of level needed to completely burn fuel Delta T Differential temperature Losses Losses calculated from oxygen and type of fuel I...

Page 3: ...igate DOWN SEND ENTER to Select and Print Particle Filter Inside Water Trap Water Trap Protective Rubber Boot w built in Magnets Battery Charge AC Adapter Flue Gas Inlet Connection Water Trap Drain 2 Line Backlit Display On Off Button Rotary Test Selector Dial Battery Compartments in back under boot AC Power Charge Indicator Light Temperature Connections Flue Probe Temp T1 Flue Probe Temperature P...

Page 4: ...nce all digits are set Press ENTER to complete setting the date once all digits are set C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time MENU SET TIME C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time SET TIME 08 30 15 _ C85 L O N G L I F E E O S A N A L ...

Page 5: ...O Temp Fuel Date Time AUXPAGE1 EXIT C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time AUX PAGE2 C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time AUX EXIT AUXPAGE1 LINE 2 shows press Send to continue to set line 2 Press s UP or t DOWN select between availa...

Page 6: ...CO Temp Fuel Date Time HEADER HEADER2 C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time _PHONE NUMBER C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time HEADER EXIT The existing HEADER 1 will display with the cursor underneath the first digit Press s UP or ...

Page 7: ...E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time DEL ALL YES PRINT SETUP Rotate dial to Menu Rotate dial to Menu Press s UP or t DOWN to REPORT Press ENTER Press s UP or t DOWN to PRINTING Press ENTER Press s UP or t DOWN to DEL ALL Press ENTER to proceed Press s UP or t DOWN to select printer Press ENTER to select Analyzer will automatically return to the menu P...

Page 8: ... is approximately at the center of the stack 4 stack adjust cone to aprox 2 from end of probe Rotate selector to desired test Rotate selector to Status The status menu shows battery life and the number of days until suggested re certification CAL Rotate selector to Aux The auxilliary menu shows user defined parameters The display alternates between page 1 and page 2 To set up the auxilliary menu s...

Page 9: ...Free The O2 CO Air free screen shows the calculated amount of O2 from CO2 measured and fuel selected and CO Air Free CO Air free is a calculated value for carbon monoxide as if there were no dilution from air COppm 20 9 20 9 O2 measured Rotate test selector to Fuel The fuel selector position will display the current fuel being used for calculations Press s UP or t DOWN to scroll and press ENTER to...

Page 10: ...tion O2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Send and Return Water temp 80 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas Pressure Test Limit Switch Pressure Switch Verify proper operation Static Duct Pressure Temperature Rise AC side Static Pressure Drop across coils 90 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Effic...

Page 11: ...o 9 O2 Stack Temperature 325 to 500 F Draft Water Column Inches 02 to 04wc Carbon Monoxide parts per million 100ppm Gas Fired Power Burners Oxygen 3 to 6 O2 Stack Temperature 275 to 500 F Stack Draft Water Column Inches 02 to 04wc Overfire Draft Water Column Inches 02wc Carbon Monoxide parts per million 100ppm Oil Fired Burners 2 Oil Fuel Oxygen 4 to 7 O2 Stack Temperature 325 to 600 F Stack Draft...

Page 12: ...ter MUST be changed when it appears discolored on the inner surface Remove water trap assembly from the analyzer as shown above Remove the filter and plastic holder from the housing Discard the filter element but keep the holder to fit to the new filter Clean the inside of the filter housing with a suitable soft cloth Fit the holder onto the new filter element and then insert into the housing Refi...

Page 13: ...stent with local disposal regulations WARNING Repair and service of this instrument is to be performed by qualified personnel only Improper repair or service could result in physical degradation of the instrument This could alter the protection from personal injury this meter provides to the operator Perform only those maintenance tasks that you are qualified to do Remove the probe from the flue a...

Page 14: ... and the stack tempera ture and then reference a slide scale that would give you the relative efficiency number UEi s electronic combustion analyzers perform the measurements on a continuous basis and can calculate the efficiency as adjustments are being made Combine this with a printout and you are able to provide a before and after comparison of the combustion equipment in relatively little time...

Page 15: ...s certified to be compliant The European Council Directive 89 336 EEC requires that electronic equipment does not generate electromagnetic disturbances that exceed defined levels and has an adequate level of immunity to enable it to be operated as intended Since there are many electrical products in use that pre date this Directive and may emit electromagnetic radiation in excess of the standards ...

Page 16: ...Canadian customers we have a recertification service center in Vancouver BC which is ISO IEC 17025 2005 accredited and offers NIST certification services for select products All recertification services are warranted for one year Save time with online recertification check in by visiting www ueitest com and click the SERVICE REQUEST banner and filling out the form Our current recertification price...

Reviews: