Enter each digit of the code as explained in Set Up Menu (Section 5.2.5). Once the last digit has been entered
the factory settings will be restored and the analyzer will return to the main display. A calibration check
should now be performed.
If the wrong code is entered the analyzer will exit to the previous menu without restoring the factory settings.
11. PRODUCT SPECIFICATION FOR AUTO2-2
*1
Using dry gases at STP
To obtain the quoted specification an instrument should be calibrated with clean ambient air (normally outside
the workshop) at standard temperature and pressure (STP).
Note:
The analyzer is not for use with a Dynamometer, unless the high temperature probe is fitted.
NDIR means by non-dispersive infrared principle.
Parameter
Resolution
Accuracy
Range
Carbon Monoxide
0.01 %
+/- 10 % of reading
*1
0-10 %
(CO)(Infrared - NDIR)
+/- 0.2 % volume
*1
Over-range 20 %
Hydrocarbon
1 ppm
+/- 10 % of reading
*1
0-5000 ppm
(HC - hexane)(NDIR)
+/- 30 ppm volume
*1
Over-range: 10,000 ppm
CO & HC
OIML R99 Class 2
Oil Temperature
0.1˚C/F
±2.0˚C ±0.3% or reading
0-150˚C
±3.6˚F ±0.3% of reading
32-302˚F
RPM
1 rpm
50 rpm
200-6,000 rpm. 2 or 4 stroke engines.
Inductive clamp or low tension connections.
Additional features
Specifications
Sensor response T
95
Nominal response time
of 20 seconds
Warm up
Less than 3 minutes
Pre-programmed Fuels
Gasoline, LPG and CNG
Data-Logging
255 Snap Shots
PC Connections
Via RS 232 port
Dimensions
Weight
1Kg
Handset
220mm x 55mm x 120mm
Probe
Insertion depth 350mm x Diameter 15mm
Clip handle to secure to exhaust, 4m long hose
Various probes available including high temperature
Ambient Operating Range
+5˚C to 45˚C/10% to 90% RH non condensing
Storage temperature
Minimum: 0˚C
Maximum: +50˚C
Battery Charger
Input: 110 Vac/230 Vac nominal
Output: 10 Vac off load
Analyzer battery run time
>4 hours from full charge with the pump running