REPLACING THE BATTERIES

This meter has been designed for use with both alkaline and rechargeable Nickel Metal Hydride (NiMH) batteries. No other types are 

recommended and will void warranty. The analyzer is supplied with 3 (AA) size NiMH rechargeable batteries. These should be installed 

into the instrument.

CAUTION

Take great care when installing the batteries to observe correct polarity.    

Always check the meter for operation immediately after installing new batteries.

Using Rechargeable Batteries

The battery charger must only be used when NiMH batteries are fitted. Alkaline batteries are not rechargeable. Attempting to recharge 

alkaline batteries may result in damage to the product and create a fire risk.

WARNING

Do not try and charge the unit with Alkaline batteries fitted. Do not mix NiMH cells of different capacities of from  

different manufacturers, all cells must be identical.

Turn over the analyzer, remove the protective rubber boot and fit 3 “AA” NiMH rechargeable batteries in the battery compartment. 

Take 

great care to ensure they are fitted with the correct battery polarity. 

Replace the battery cover and the protective rubber boot.

Time and Date

When changing the batteries, the time and date will need setting.

Battery Disposal 

Always dispose of depleted batteries using approved disposal methods that protect the environment.

Battery Charging

Ensure that you use the correct charger. This unit uses a 5V regulated charger. Ensure the batteries are fitted in the correct manner, 

and charge for at least 16 hours. Subsequent charges should be overnight. NiMH batteries may be charged at any time, even for short 

periods to conduct testing.

WARNING

Under NO circumstances should you expose batteries to extreme heat or fire as they may explode and cause injury. 

Always dispose of old batteries promptly in a manner consistent with local disposal regulations.

ANNUAL SERVICE & RECERTIFICATION

What You Get

Calibration (or sensor replacement)
30-point inspection by factory trained professionals
Firmware and software updates
Certificate of calibration
Another 12-month warranty on the analyzer
Return unit freight paid – within continental US and Canada
List of work performed, and parts replaced

Summary of Contents for C161

Page 1: ...547 5740 Fax 503 643 6322 www ueitest com Email info ueitest com INSTRUCTION MANUAL ENGLISH Flue Gas Combustion Analyzer STATUS F L U E 1 AUX FLUE 2 T E M P Combustion Analyzer C161 CO C O 2 T E M P Type text here ...

Page 2: ...con Layout 10 Status Bar Menu Options 10 Status Bar Options Table 10 Menu 11 Using the Menu 11 Stored Logs Memory 12 Menu Options 12 Viewing Stored Logs 13 Log View Menu Options 13 Navigating Stored Logs 13 Log Navigation Menu Options 14 Temperature Testing 14 Temperature Display 14 Viewing Printing 14 Printouts 15 Specifications 16 Certification 16 Where to Test 17 18 What Results are Generally C...

Page 3: ...s who are familiar with the hazards of their trade Observe all recommended safety procedures that include proper lockout utilization and use of personal protective equipment that includes safety glasses gloves and flame resistant clothing Users of portable gas detectors are recommended to conduct a bump check before relying on the unit to verify an atmosphere is free from hazard A bump test is a m...

Page 4: ...Module is fitted K Down Button Short press to navigate DOWN L ENTER Button Short press select current option displayed Long press to activate some menu items M Rotary Selector Dial N Particle Filter inside water trap O Water Trap P LED Full Water Trap Indicator Q Serial Number under Protective Boot R Sensors Fitted under Protective Boot Indicates Sensors fitted in unit S Battery Compartment under ...

Page 5: ...n Analyzer C161 CO C O 2 T E M P Serial No Sensors Fitted CO CO2 NO O2 T2 T1 B C D E F G H I J K L M W V X P O N S U T R Q Flue Probe Temperature Plug Plugs into T1 Narrow Pin MUST be on the Right hand side Flue Probe Gas Inlet Plug Black Connector plugs into Flue Gas Inlet ...

Page 6: ...essure Differential Pressure NO Nitric Oxide Sensor Wireless Module Over Range Pump PRE TEST CHECKLIST SETTING INLET TEMPERATURE ANALYZER CONNECTIONS Turn on and zero the analyzer without the flue probe connected to use ambient temperature Connect flue probe thermocouple to T1 during zero countdown use for inlet ducted system Connect a thermocouple to T2 to measure second temp source Clean particl...

Page 7: ...to CO2 Zero On the Purge option short press the Edit Button and a 60 second Purge will begin STATUS F L U E 1 AUX FLUE 2 T E M P P EMPTYING CLEANING THE IN LINE WATER TRAP CHANGING THE PARTICLE FILTER QUICK START Turn on the analyzer by pressing the On Off Button for 2 seconds until the unit activates As described the analyzer will perform a 60 second turn on purge Once completed simply select the...

Page 8: ...Efgc except for Heavy Oil and Pellets O2 Ref Selectable parameter select between 0 0 to 20 0 Logs if logs are saved they will be displayed Code for Authorized Service Personnel only Status Screen Ti Ambient temperature ATM atmospheric pressure in mBars CAL days until next calibration due AUX Screen Default selections can be changed by user to desired selections select between O2 CO Display Fuel Ty...

Page 9: ...PG Butane Propane Natural Gas Bio Oil Ambient temperature around the analyzer Current atmospheric pressure mBar Shows number of days until next calibration is due Status Bar User Interface The large six line display will indicate 5 lines and the status bar see pg 9 the backlight will activate on each button press and turn off after 10 seconds of no activity Navigation through the various start opt...

Page 10: ...e Status Bar will offer the user Menu items based on the Selector Dial setting Status Bar Options Table Test Screen Dial Digital Position Test Screen Status Info Available STATUS AUX FLUE 1 FLUE 2 TEMP TIME 4 4 4 4 4 FLUE SELECTED 4 4 4 4 4 DATE 4 4 4 4 4 CO2 FLUE 4 4 4 Date Option Date Displayed as Pump Pump Sending Data Battery Pump on Pump paused Indicates data is being transmitted to either th...

Page 11: ...Press and buttons to change field contents Press to enter content selected Menu Items MENU Sub Menu OPTIONS COMMENTS Time HH MM SS 24 Hour Format e g 7AM 0 00 00 7PM 19 00 00 Date DD MM YY Header LINE 1 Customer info displays on print outs LINE 2 Customer info displays on print outs EXIT IR Print KMIRP Select to print to KMIRP IRP 2 Select to print to IRP 2 Temperature Scale C Secet F or C F Gas U...

Page 12: ...usage status simply navigate to menu LOGS press the button to select the Log desired Number of stored Combustion Logs Number of stored Auxiliary Logs Number of stored Temperature Logs and Pressure if fitted Total Memory Used Menu Options Menu Options STATUS F L U E 1 AUX FLUE 2 T E M P Combustion Analyzer C161 CO C O 2 T E M P View stored Logs this will enter the secondary menu to select the type ...

Page 13: ...ill be displayed Log Number Log time and date this line will alternate between time and date of Log Log readings information is specific to Log type Navigation Menu options STATUS F L U E 1 AUX FLUE 2 T E M P Combustion Analyzer C161 CO C O 2 T E M P View stored Combustion Logs Option will appear if Logs are available View stored Auxiliary Logs Option will appear if Logs are available Number of st...

Page 14: ...g of the selected type Option will only appear if there is more than one Log Navigate to the previously selected Log of the selected type Option will only appear if there is more than one Log Exit back to the previous menu STATUS F L U E 1 AUX FLUE 2 T E M P T E M P Pressure will only show if fitted Use the T1 connection for the Supply temperature sensor Use the T2 connection for the Return temper...

Page 15: ...ting can be aborted by pressing the PRINT BUTTON again UEi Infrared Printer Make sure the printer is switched on ready to accept data and its infrared receiver is in line with the analyzer s infrared emitter on top of the analyzer Combustion UEi Test Instruments SW20157 0 3083 0 11 20 76 1 0 0009 83 9 T1 T2 Ti NETT 11 16 19 09 26 20 Auxiliary 0 1 11 16 19 09 26 20 UEi Test Instruments SW20157 0 30...

Page 16: ...50379 Parts 1 3 in accordance to 1st German Feder al Emission Control Ordinance Bim5chV Operating Conditions Temperatures Humidity 32 to 113 F 0 to 45 C 15 to 90 RH non condensing Power Supply Rechargeable batteries USB Charging Physical Characteristics Weight Dimensions 1 5 lb 635g H 8 54 x W 4 18 x D 1 86 H 217 mm x W 106 mm x D 47mm The C161 is in conformity with the relevant Union harmonizatio...

Page 17: ...ers Draft Verify proper combustion O2 CO2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Supply and Return Water temp 80 Furnace Verify proper combustion O2 CO2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas Pressure Test Limit Switch Pressure Switch Verify proper operation Static Duct Pressure Temperature Rise AC side Static Pressure Drop across coils 90 Furnace Verify proper combust...

Page 18: ...Static Duct Pressure Temperature Rise AC side Static Pressure Drop across coils WHAT RESULTS ARE GENERALLY ACCEPTABLE What Results Are Generally Acceptable Atmospheric Gas Fixed Burners Gas Fired Burners Oil Fired burners 2 Oil Fuel Positive Overfire Gas Oil Oxygen 7 to 9 O2 3 to 6 O2 4 to 7 O2 3 to 9 O2 Stack temperature 325 F to 500 F 275 F to 500 F 325 F to 600 F Draft Water Column Inches 02 to...

Page 19: ...ds can degrade electronic circuits Whenever practical keep the instrument away from dust and dirt that can cause premature wear Although your instrument is built to withstand the rigors of daily use it can be damaged by severe impacts Use reasonable caution when using and storing the meter COLD WEATHER PRECAUTIONS It is important you keep your flue gas analyzer in a warm and dry place overnight El...

Page 20: ...ery compartment Take great care to ensure they are fitted with the correct battery polarity Replace the battery cover and the protective rubber boot Time and Date When changing the batteries the time and date will need setting Battery Disposal Always dispose of depleted batteries using approved disposal methods that protect the environment Battery Charging Ensure that you use the correct charger T...

Page 21: ...ecked Packing Your Analyzer Put analyzer and probes back in their cases The case should be put into a box with 1 2 inches of packing on each side for protection When shipping an analyzer only use a shoe box size container with enough packing to fill the empty space Print out the service paperwork sent to you and include it in the packaging NOTE If you are having specific problems that you want eva...

Page 22: ... measure the CO2 level and the stack temperature and then reference a slide scale that would give you the relative efficiency number UEi s electronic combustion analyzers perform the measurements on a continuous basis and can calculate the efficiency as adjustments are being made Combine this with a printout and you are able to provide a before and after comparison of the combustion equipment in r...

Page 23: ......

Page 24: ...aired or replaced at UEi s option This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration tampering accident misuse abuse neglect or improper maintenance Batteries and consequential damage resulting from failed batteries are not covered by warranty Any implied warranties including but not limited to implied warranties of merchanta...

Reviews: