• Remove the rubber plug
• Allow the water to drain out
• Re-insert the rubber plug
• Remove the protective rubber boot
• Slide the water trap unit from the analyzer
• Remove the particle filter from its’ spigot and replace
• Slide the water trap back into position and replace protective rubber boot
FRESH AIR PURGE
Each time the analyzer is turned on it will perform a 60 second air purge, this is to clear
the gas sampling path (including probe, if connected), and self calibrate the CO
2
detection system. For these reasons it is very important that the analyzer be in
outside
fresh air
when powered on.
WARNING
When in the Rotary Dial positions AUX, Flue 1 or Flue 2 ; Press the Up or Down Buttons,
scroll to CO2 Zero. On the Purge option, short press the Edit Button, and a 60 second
Purge will begin.
STATUS
FLUE 1
AUX
FLUE 2
TEMP
Combu
stion Analy
zer
C161
CO
C
O2
T
EM
P
EMPTYING & CLEANING THE IN-LINE WATER TRAP
CHANGING THE PARTICLE FILTER
QUICK START
Turn on the analyzer by pressing the On/Off Button for 2 seconds until the unit activates. As described the analyzer will perform a 60
second turn on purge. Once completed simply select the reading desired by the position of the rotary dial.
MEASURING FLUE GASES
After the initial countdown is finished and the analyzer is properly setup, put its’ flue probe in the appliance’s sampling point. The tip of
the probe should be at the center of the flue. Use the flue probes depth stop cone to set the position.
With balanced flues, make sure the probe is positioned into the flue so no air can “back flush” into the probe.
NOTE:
Ensure the flue probe handle does not get hot!
20”
Boiler
Flue