background image

C

o m b u s t i o n

  m

a d e

  e

a s y

Instruction Manual

C20 

Combustion Set-Up 

Meter

99 Washington Street 
Melrose, MA 02176 

Phone

 781-665-

1400

Toll Free 1-800-517-8431

Visit us at www.T

estEquipmentDepot.com

Summary of Contents for C20

Page 1: ... o m b u s t i o n M a d e E a s y Instruction Manual C20 Combustion Set Up Meter 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Page 2: ...1 Overview 2 Maintenance General Maintenance 3 Water Trap 3 Changing Particle Filter 3 Battery Replacement 4 Electromagnetic Compatibility 4 Combustion Information Why Test With The C20 5 Acceptable Results 6 Meter Problem Solving 7 Glossary Selector and Display Parameters 8 Meter Specifications 9 Warranty 10 ...

Page 3: ...n ON the pump Always allow the readings to return to zero 00 for CO prior to shutting the unit off WARNING The probe will be hot from flue gases Remove the probe from the flue and allow it to cool naturally Do not immerse the probe in water as this will be drawn into the analyzer and damage the pump and sen sors Once the probe is removed from the flue and the readings have returned to ambient leve...

Page 4: ... 0 0 CO to Zero Be careful as the probe tip will be HOT Drain water trap Check particle filter Rotate selector to OFF NOTE To protect the sensor HIGH will be displayed and the zero countdown will repeat until CO measured is less than 20 ppm C20 Overview OFF S t a n d b y C O CO2 COMax C F 0000 LCD Display Selector Dial Water Trap ...

Page 5: ...ents or strong detergents as they may damage the finish impair safety or affect the reliability of the structural components Emptying Cleaning the In Line Water Trap The in line water trap should be checked and emptied on a regular basis Water vapor will condense in the probe line which may cause the water trap to fill suddenly if the probe is moved Care should be taken at all times Carefully un s...

Page 6: ...e operated as intended Since there are many electrical products in use that pre date this Directive and may emit electromagnetic radiation in excess of the standards defined in the Directive there may be occasions where it would be appro priate to check the analyzer prior to use The following procedure should be adopted Go through the normal start up sequence in the location where the equipment is...

Page 7: ... gases could produce carbon monoxide in deadly quantities The UEi C20 can be used to test both flue gas and ambient CO When customers complain about fumes it is usually an indication of improper operation CO is a colorless odorless gas so the fumes are not the CO but an indication that a problem may exist Improve Your Productivity Profit The UEi C20 combustion set up meter gives a quick continuous...

Page 8: ...8 0 CO2 Stack Temperature 325 to 600 F Stack Draft Water Column Inches 04 to 06wc Overfire Draft Water Column Inches 02wc Carbon Monoxide parts per million 100ppm Smoke 0 or manufacturer s recommendation Positive Overfire Gas Oil CO2 5 8 to 10 3 CO2 Stack Draft Water Column Inches 02 to 04wc Overfire Draft Water Column Inches 0 4 to 0 6wc Carbon Monoxide parts per million 100ppm NOTE Follow manufa...

Page 9: ...O or CO2 too low Air leaking into probe tubing water trap or connectors CO or CO2 reading Meter was stored in a cold environment and is not at normal working temperature Batteries not holding charge Batteries exhausted Meter does not respond to flue gas Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminants 7 ...

Page 10: ...million ppm COMAX Highest measured value of Carbon Monoxide since the instrument was powered on CO2 Carbon Dioxide Direct reading of the carbon dioxide sensor displayed in percentage BAT Displays the Battery power available HIGH Displays at standby power off if CO levels are greater than 20ppm NOTE If turning unit off countdown will resume once CO levels are lesser than 20ppm GLOSSARY ...

Page 11: ...lb Handset 7 x 2 x 31 4 Probe 4 x 1 4 stainless steel shaft nominal 36 neoprene hose Ambient Operating Range 32 F to 104 F 10 to 90 RH non condensing 850 to 1100 mmHg atmospheric pressure Battery Life 4 AA cells 8 hours using Alkaline AA cells 1 Using dry gases at STP 2 Calculated 3 Calculated assuming fuel lean combustion METER SPECIFICATIONS NOTE MAY BE SUBJECT TO CHANGE ...

Page 12: ...ties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or eco nomic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of original purchase date will be required before warranty repairs wil...

Reviews: