ELECTROMAGNETIC COMPATIBILITY (EMC)
This product has been tested for compliance with the following generic standards: EN 50081-1, EN 50082-1 and is certified to be compliant .
The European Council Directive 89/336/EEC requires that electronic equipment does not generate electromagnetic disturbances that exceed defined levels and
has an adequate level of immunity to enable it to be operated as intended .
Since there are many electrical products in use that pre-date this Directive and may emit electromagnetic radiation in excess of the standards defined in the
Directive there may be occasions where it would be appropriate to check the analyzer prior to use . The following procedure should be adopted .
• Go through the normal start up sequence in the location where the equipment is to be used
• Switch on all localized electrical equipment that might be capable of causing interference
• Check that all readings are as expected (a level of disturbance in the readings is acceptable)
• If not, adjust the position of the instrument to minimize interference or switch off, if possible, the offending equipment for the duration of the test
At the time of writing this manual (April 2015) UEi is not aware of any field based situation where such interference has ever occurred and this advice is only
given to satisfy the requirements of the Directive .
SPECIFICATIONS
Parameter
Range
Resolution
Accuracy
Temperature Measurements
Flue Temperature
32 - 1112˚F (0 - 600˚C)
1 .0˚ F˚/C˚
±(0 .3% 3 .6˚F(2 .0˚C))
Inlet Temperature (Internal sensor)
32 ~ 122 ˚F (0-50˚C)
1 .0˚ F˚/C˚
±(0 .3% 1 .8˚F(1 .0˚C))
Inlet Temperature (External sensor)
32 - 1112˚F (0 - 600˚C)
1 .0˚ F˚/C˚
±(0 .3% 3 .6˚F(2 .0˚C))
Gas Measurements
Carbon Monoxide (CO)
*1
0 - 60 ppm
1ppm
±3ppm
61 - 4,000ppm nom
1ppm
±5% of reading
4,001ppm to 10,000ppm
1ppm
±10% of reading
20,000ppm max for 15 min
1ppm
Not specified
Carbon Dioxide (CO2)
*1
0 - 20%
0 .1%
±0 .3%
Calculations
Oxygen (O2)
*2
0 - 21%
0 .1%
±0 .3%
Efficiency
*2
0 - 99 .9%
0 .1%
±1 .0% of reading
Excess Air
*2
0 - 250%
0 .1%
±0 .2% of reading
CO/CO2 ratio
*2
0 - 0 .999
0 .0001
±5% of reading
Nitric Oxide Measurement (C257)
Nitric Oxide (NO) low
2 - 30 ppm
1 ppm
±2 ppm < 30ppm
31 - 100 ppm
1 ppm
±5 ppm > 30ppm
Nitric Oxide (NO) high
5 - 100 ppm
1 ppm
±5 ppm < 100ppm
101 - 1000 ppm
1 ppm
±5% reading >100ppm
Pressure
Nominal range ±64 .2 InH2O (160mBar) Max
over-range w/o damage to sensor is ±321In-
H2O (800 mBar)
±0 .6 inH2O
Maximum 0 .001 InH2O
<10 InH2O
±0 .02 In H2O
±64 .2 inH2O
±3% of reading
Pre-programmed Fuels
Natural gas, L Oil, H Oil, Propane, Wood, Biofuel, Wood, PLUS 3 User defined fuels .
Storage Capacity
99 Combustion tests, 20 Pressure tests, 20 Heat Exchange tests, 20 Temperature tests
20 Room CO tests
GENERAL SPECIFICATIONS
• Operating Altitude: 2000m (6,561 ft.)
• Storage Altitude:
10,000m (32,808 ft .)
• Operating Temperature:
32°F to 104°F (0°C to 40°C) at 10 - 90% R .H
• Pollution Degree:
2
• Dimensions:
Handset: 7 .9” (200mm) x 3 .5” (90mm) x 1 .8” (45mm)
Probe: (L) 11 .8” (300mm) x (D) 0 .25” (6mm) with 7 .8” (200mm)
6ft (3m) neoprene hose
• Weight:
2 .2lbs (1kg)
• Certifications:
EMC
EN 50081-1, EN 50082-1, CE
• Battery Type:
4 x 1 .5V AA
• Battery Life:
>8Hrs using AA Alkaline
• Power Supply Input:
110V AC Output: 9V DC Regulated