background image

19

ELECTROMAGNETIC COMPATIBILITY (EMC)

This product has been tested for compliance with the following generic standards:  EN 50081-1, EN 50082-1 and is certified to be compliant .

The European Council Directive 89/336/EEC requires that electronic equipment does not generate electromagnetic disturbances that exceed defined levels and 
has an adequate level of immunity to enable it to be operated as intended .

Since there are many electrical products in use that pre-date this Directive and may emit electromagnetic radiation in excess of the standards defined in the 
Directive there may be occasions where it would be appropriate to check the analyzer prior to use . The following procedure should be adopted .

• Go through the normal start up sequence in the location where the equipment is to be used
• Switch on all localized electrical equipment that might be capable of causing interference
• Check that all readings are as expected (a level of disturbance in the readings is acceptable)
• If not, adjust the position of the instrument to minimize interference or switch off, if possible, the offending equipment for the duration of the test

At the time of writing this manual (April 2015) UEi is not aware of any field based situation where such interference has ever occurred and this advice is only 
given to satisfy the requirements of the Directive .

SPECIFICATIONS

Parameter

Range

Resolution 

Accuracy

Temperature Measurements

Flue Temperature

32 - 1112˚F (0 - 600˚C) 

1 .0˚ F˚/C˚

±(0 .3% 3 .6˚F(2 .0˚C))

Inlet Temperature (Internal sensor)

32 ~ 122 ˚F  (0-50˚C)

1 .0˚ F˚/C˚

±(0 .3% 1 .8˚F(1 .0˚C))

Inlet Temperature (External sensor)

32 - 1112˚F (0 - 600˚C)

1 .0˚ F˚/C˚

±(0 .3% 3 .6˚F(2 .0˚C))

Gas Measurements

Carbon Monoxide (CO)

*1

0 - 60 ppm

1ppm

±3ppm

61 - 4,000ppm nom

1ppm

±5% of reading

4,001ppm  to 10,000ppm

1ppm

±10% of reading

20,000ppm max for 15 min

1ppm

Not specified

Carbon Dioxide (CO2)

*1

0 - 20%

0 .1% 

±0 .3% 

Calculations

Oxygen (O2)

*2

0 - 21%

0 .1%

±0 .3%

Efficiency

*2

0 - 99 .9%

0 .1%

±1 .0% of reading

Excess Air

*2

0 - 250%

0 .1%

±0 .2% of reading

CO/CO2  ratio

*2

0 - 0 .999

0 .0001

±5% of reading

Nitric Oxide Measurement (C257)

Nitric Oxide (NO) low

2 - 30 ppm

1 ppm

±2 ppm < 30ppm

31 - 100 ppm

1 ppm

±5 ppm > 30ppm

Nitric Oxide (NO) high

5 - 100 ppm

1 ppm

±5 ppm < 100ppm

101 - 1000 ppm

1 ppm

±5% reading >100ppm

Pressure

Nominal range ±64 .2 InH2O (160mBar) Max 

over-range w/o damage to sensor is ±321In-

H2O (800 mBar)

±0 .6 inH2O

Maximum 0 .001 InH2O 

<10 InH2O

±0 .02 In H2O

±64 .2 inH2O

±3% of reading

Pre-programmed Fuels

Natural gas, L Oil, H Oil, Propane, Wood, Biofuel, Wood, PLUS 3 User defined fuels .

Storage Capacity

99 Combustion tests, 20 Pressure tests, 20 Heat Exchange tests, 20 Temperature tests

20 Room CO tests

GENERAL SPECIFICATIONS

•  Operating Altitude:  2000m (6,561 ft.) 

•  Storage Altitude:

 10,000m (32,808 ft .)

•  Operating Temperature:

 32°F to 104°F (0°C to 40°C) at 10 - 90% R .H

•  Pollution Degree:

 2

•  Dimensions:

 Handset: 7 .9” (200mm) x 3 .5” (90mm)  x 1 .8” (45mm) 

  Probe: (L) 11 .8” (300mm) x (D) 0 .25” (6mm) with 7 .8” (200mm)   

 

  6ft (3m) neoprene hose

•  Weight:

 2 .2lbs (1kg)

•  Certifications:

 

EMC

 

EN 50081-1, EN 50082-1, CE

•  Battery Type:

 4 x 1 .5V AA

•  Battery Life:

 >8Hrs using AA Alkaline

•  Power Supply Input:

 110V AC Output: 9V DC Regulated

www.

.com

information@itm.com

1.800.561.8187

Summary of Contents for C257

Page 1: ...nter I N D U S T R I A L E O S A N A L Y Z E R CO2 H2CO T PRS O2 9 7 COp 28 P 1 47 TFf 87 4 C255 C257 1 800 547 5740 Fax 503 643 6322 www ueitest com Email info ueitest com INSTRUCTION MANUAL ENGLISH...

Page 2: ...n Measurement Terms 14 Specifications 15 Electromagnetic Compatibility 15 Limited Warranty 16 DISPLAY SYMBOLS Excess Air Theoretical amount of air in excess of level needed to completely burn fuel Eff...

Page 3: ...sure SEND ENTER to Select and Print press and hold to log current values Particle Filter Inside Water Trap Water Trap Protective Rubber Boot w built in Magnets Battery Charge AC Adapter Flue Gas Inlet...

Page 4: ...O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter MENU SETUP C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status A...

Page 5: ...P Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter LANGUAGE ENGLISH C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room...

Page 6: ...N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter RESET GAS ZERO IN FRESH AIR C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEM...

Page 7: ...CH Heat exchange integrity test TEMP and ROOM CO tests Press ENTER to view Press s UP or t DOWN to select VIEW DEL ALL or EXIT Select View to display logs Select Del All to erase records or Exit to mo...

Page 8: ...A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter AUX LINE 1 NOp_ N F C225 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Fuel Flue Te...

Page 9: ...ck adjust cone to aprox 2 from end of probe Rotate dial to Flue Test Rotate dial to Status The status menu shows battery life time date and the number of days until suggested re certification CAL Rota...

Page 10: ...Room CO Up Down Enter I N D U S T R I A L E O S A N A L Y Z E R CO2 H2CO T PRS O2 9 7 COp 28 P 1 47 TFf 87 4 Menu Status Aux Fuel Flue Test Prs Exch Test Temp Room CO Up Down Enter C225 I N D U S T R...

Page 11: ...rning on and after readings have stabilized press the SEND button This will store the Pre Blower test segment After the blower turns on press PUMP to start the Post Blower test The meter will wait 60...

Page 12: ...page 7 Press the PUMP button to start the test CO readings peak will be logged every 2 minutes for a 30 minute time span To see logged results see page 7 Menu Status Aux Fuel Flue Test Prs Exch Test T...

Page 13: ...E after a 4 second countdown and automatically zero the pressure sensor Analyzer model number time date and countdown will show C225 L O N G L I F E E O S A N A LY Z E R CO2 CO TEMP Menu Status Aux Fu...

Page 14: ...O2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Supply and Return Water temp 80 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas Pressure Test Limit S...

Page 15: ...9 O2 Stack Temperature 325 to 500 F Draft Water Column Inches 02 to 04wc Carbon Monoxide parts per million 100ppm Gas Fired Power Burners Oxygen 3 to 6 O2 Stack Temperature 275 to 500 F Stack Draft W...

Page 16: ...the pump and sensors that will cause damage The filter MUST be changed when it appears discolored on the inner surface Remove water trap assembly from the analyzer as shown above Remove the filter an...

Page 17: ...local disposal regulations WARNING Repair and service of this instrument is to be performed by qualified personnel only Improper repair or service could result in physical degradation of the instrumen...

Page 18: ...nd the stack tempera ture and then reference a slide scale that would give you the relative efficiency number UEi s electronic combustion analyzers perform the measurements on a continuous basis and c...

Page 19: ...C 0 3 reading 3 6 F 2 0 C Inlet Temperature Internal sensor 32 122 F 0 50 C 1 0 F C 0 3 reading 1 8 F 1 0 C Inlet Temperature External sensor 32 1112 F 0 600 C 1 0 F C 0 3 reading 3 6 F 2 0 C Gas Mea...

Page 20: ...products All recertification services are warranted for one year Save time with online recertification check in by visiting www ueitest com and click the SERVICE REQUEST banner and filling out the for...

Reviews: