background image

Replacement Part s

Battery

AB10

Replacement Sensor

RS1000

Replacement Sensor Cap

RS1010

Optional Accessories

Headset Earphones

HE1000

Hard Carrying Case

AC501

E l e c t romagnetic Compatibility

The European Council Directive 89/366/EEC requires that electronic
equipment does not generate electromagnetic disturbances that exceed
defined levels and has an adequate level of immunity to enable it to be
operated as intended. The specific standards applicable to this product
are detailed in the appendices.

Since there are many electrical products in use that pre-date 
this Directive and may emit electromagnetic radiation in excess 
of the standards defined in the Directive, it may be appropriate 
to check the analyzer prior to use. The following procedure should 
be adopted:

1.  Go through the normal start sequence in the location where 

the equipment is to be used.

2.  Switch on all localized electrical equipment that might 

cause interference.

3.  Check that all readings are as expected. (Generally, a low

disturbance is acceptable.) If not, adjust the position of 
the instrument to minimize interference or switch off the 
offending equipment during the test. There has been no 
awareness of any field based situation where such interference
has ever occurred since the writing of this manual (April 
1998). This advice is only given to satisfy the requirements of 
the Directive.

This product has been tested for compliance with the following generic
standards, and it is certified to be compliant.

EN 50081-1
EN 50082-1

Specifications EC/EMC/K1/CD100A details the specific test 
configuration, performance and conditions of use.

B a t t e ry Replacement

1.  Turn off the power and remove the battery case cover.

2.  Remove and dispose of old 9V alkaline battery promptly in 

a manner consistent with disposal regulations. Always use 
a 9V alkaline battery. Ensure the polarity of the battery matches
that shown in the compartment.

3.  Replace the battery case cover.

Sensor Filter and Sensor Replacement

The sensor is housed at the tip of the gooseneck assembly, and 
has a life of more than three years.

To remove the sensor:

1.  Turn the unit off.

2.  Pull off the top half of the sensor housing. (Pushing to 

one side aids removal)

3.  With care, move the "Tip-Light" LED to one side to gain 

access to the sensor (small silver can).

4.  Unplug the sensor and replace it with a new one.

NOTE:

The sensor can be fitted in either one of two ways without

affecting instrument operation.

5.  Realign the LED, ensuring it sits above the sensor.

6.  Replace the sensor cover, by pushing it firmly in place. 

Note the locating pin between the body and sensor cap.

S p e c i f i c a t i o n s

Protection against electric shock in accordance with 
EN 61010-1: 1993

This instrument is designated as Class III, SELV.

Power supply:

Sensor:

Sensitivity:

Indicators:

Warm-up:

Response Time:

Duty Cycle:

Battery Life:

Dimensions:

Weight:

Probe Length:

O p e ra t i v e :

1 x 9V battery

Solid state semi conductor

< 50 ppm methane

Audible: Adjustable tic rate
Visual: Flashing LED

10 seconds

Instantaneous

Continuous

5 hrs. typical use

8" (H) x 4" (W) x 1/2" (D)

205mm x 100mm x 36mm

15oz. / 400mm

0 - 50˚C / +32 - 122˚F

CD100A-MAN    

P. 2

Reviews: