19
Test Lead Notes
Cat IV and CAT II Measurement Locations
• Ensure the test lead shield is pressed firmly in place. Failure to use the CAT
IV shield increases arc-flash risk.
CAT II Measurement Locations
• CAT IV shields may be removed for CAT II locations. This will allow testing
on recessed conductors such as standard wall outlets. Take care not to lose
the shields.
WARNING:
Test Lead category protections apply only to test leads and
should not be confused with the meter’s specific CAT rating. Observe the
maximum category protection indicated on the meter the test leads are
plugged into.
CAUTION
: If the test leads need to be replaced, you must use a new one
which should meet EN 61010-031 standard, rated CATIII 1000V or better.
4mm
18mm
4mm
18mm
• Rotate Battery Compartment Latches to open
position
LOCK
OPEN
• Remove battery cover
• Replace the old batteries with 2 new (AAA)
batteries
• Replace the battery cover
• Rotate Battery Compartment Latches to lock
LOCK
OPEN
Battery Replacement