DM383/DM410-MAN
P. 5
DC Current (Amps)
*DM410 only
**Plug the red test lead into the “
mAµA
” input jack for measurements
on this range
NOTE:
“
µA
” and “
mA
” ranges are protected by a 5 amp, 600 volt fuse
“
A
” range is protected by a 10 amp, 600 volt fuse
AC Current
*DM410 only
**Plug the red test lead into the “
mAµA
” input jack for measurements
on this range
NOTE:
“
µA
” and “
mA
” ranges are protected by a 5 amp, 600 volt fuse
“
A
” range is protected by a 10 amp, 600 volt fuse
Resistance (OHMS, Continuity)
Optional Accessories
Optional
Battery, 9 V . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AB9
Fuse 5A, 600 V Ceramic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AF120
Fuse 10A, 600 V Ceramic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AF125
Test lead set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ATL50
Rubber boot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AH190
Alligator clip adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AAC
Microwave diode booster leads . . . . . . . . . . . . . . . . . . . . . . . . .ATL60
Flame safeguard relay test kit . . . . . . . . . . . . . . . . . . . . . . . . . . .ATLFSG
AC/DC clamp-on adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CA2K
AC clamp-on adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CA310
Phototach adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .DPM2K
Humidity adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HM1K
Temperature adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TA2K
Function
Range
Resolution
Accuracy
*20 µ
0.01 µA
200 µ
0.1 µA
±0.5% of reading, ±1 digit
2 m
1 µA
DC A
20 m
10 µA
200 m
100 µA
±1.2% of reading, ±1 digit
**2 A
1 mA
10 A
10 mA
±2.0% of reading, ±5 digits
Function
Range
Resolution
Accuracy
*20 µ
0.01 µA
200 µ
0.1 µA
±1.0% of reading, ±3 digits
2 m
1 µA
AC A
20 m
10 µA
200 m
100 µA
±1.8% of reading, ±3 digits
**2 A
1 mA
10 A
10 mA
±3.0% of reading, ±7 digits
Function
Range
Resolution
Accuracy
200
0.1
Ω
2 K
1
Ω
Ω
20 K
10
Ω
±0.5% of reading, ±1 digit
200 K
0.1 K
Ω
2 M
1 K
Ω
20 M
10 K
Ω
±1.0% of reading, ±2 digits