background image

DRS220 User’s Guide  

 

 

10 

Tank Capacity Mode Training 

1.  While holding 

 down, press 

.  The 

SET TANK

 indicator will turn on, and 

OFF 

will show on the display.  This indicates that the Tank Capacity Mode is 

currently off. 

2.  Turn the Tank Capacity Mode on by pressing 

.  The display will show 

ON

and the Alternate Numeric Display and Percent Indicator will turn on. 

3.  Press 

 to proceed to the first step of training.  

0%

 will flash in the Alternate 

Numeric Display indicating that the weight of an empty container must be 
registered. 

4.  Apply an empty container (or a load representing the weight of an empty 

container) on the platform. Wait 5 seconds for the load to stabilize, then press 

.  

100%

 will flash in the Alternate Numeric Display indicating that the weight 

of a full container must be registered. 

5.  Apply a full container (or a load representing the weight of a full container) on the 

platform. Wait 5 seconds for the load to stabilize, then press 

.  This 

completes the training process, and the 

SET TANK

 indicator will turn off. 

6.  From this point forward, the Alternate Numeric Display will show the percent of 

material remaining in the container from 0% to 100%. 

NOTE:

 

The 100% weight must be larger than the 0% weight by at least 200g.  If not, 

Err

” will show on the Alternate Numeric Display. To clear the error, retrain as above 

using a 100% weight that is at least 200g larger than the 0% weight. During training, 
the ZERO key may be used to assist in setting the 0% and 100% values. This will 
have no effect on the training, as the Tank Capacity Mode relies on absolute, not 
relative, weights. 

 

DRS220 User’s Guide 

11 

Turning Tank Capacity Mode Off 

Perform the following steps to turn the Tank Capacity Mode Off: 

1.  While holding 

 down, press 

. The 

SET TANK

 indicator will turn on, and 

ON will show on the display. This indicates that the Tank Capacity Mode is 
currently on. 

2.  Turn the Tank Capacity Mode off by pressing 

.  The display will show 

OFF

and the Alternate Numeric Display will turn off. 

3.  Press 

 to return to normal operation. 

Turning the Tank Capacity Mode On without Re-training 

The Tank Capacity Mode can be turned back on without retraining.  The DRS220 will 
remember the previous 0% and 100% set points until the unit is re-trained as above.  To 
turn the Tank Capacity Mode on without re-training, perform the following steps: 

1.  While holding 

 down, press 

. The 

SET TANK

 indicator will turn on, and 

OFF

 will 

show on the display. This indicates that the Tank Capacity Mode is currently off. 

2.  Turn the Tank Capacity Mode on by pressing 

.  The display will show 

ON

, and the 

Alternate Numeric Display and Percent Indicator will turn on. 

3.  Press 

 to return to normal operation. 

Battery Level Indicator 

The battery level indicator shows the relative strength of the batteries.  Full power is 
indicated by four bars.  As the batteries are depleted, the number of bars displayed 
decreases, until no bars are left.  At this point, it is necessary to replace the batteries (refer 
to the Battery Installation section above).  If the battery power drops to the point where the 
DRS220 can no longer function accurately, the alarm will beep 20 times and the power will 
turn off automatically.  When the batteries are replaced, and the scale is turned back on, the 
DRS220 will resume operation with all previous setting intact. 

Summary of Contents for DRS220

Page 1: ...ies of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages e...

Page 2: ...erant charging for A C systems Features Large high visibility LCD display Measures Weight in kgs and lbs Built in memory remembers all previous settings Refrigerant Tank Capacity Indicator Rugged Comp...

Page 3: ...y B Alternate Numeric Display C Tank Capacity Percent Indication D Units Indicators E Sound Mute Indicator F Floating Negative Indicator Indicates weight removed from platform G Battery Level Indicato...

Page 4: ...ment units by repeated pressing The units will cycle as follows kilograms pounds pounds and ounces ounces 6 Select the desired display resolution by repeatedly pressing as described in the RESOLUTION...

Page 5: ...ld set of batteries 3 Insert 4 AAA Replace battery cover by aligning tabs and snapping back into place IMPORTANT TO PREVENT DAMAGE TO THE DRS220 CAREFULLY OBSERVE BATTERY POLARITY AS MARKED INSIDE THE...

Page 6: ...display DRS220 User s Guide 9 4 Press and or hold the or keys to set the desired alarm set point To change the set point faster press This will activate the FAST SET mode which will change the set poi...

Page 7: ...on absolute not relative weights DRS220 User s Guide 11 Turning Tank Capacity Mode Off Perform the following steps to turn the Tank Capacity Mode Off 1 While holding down press The SET TANK indicator...

Page 8: ...pply the calibration weight to the platform A number representing the applied reference weight will appear on the display Allow at least 10 seconds for this number to stabilize 8 Press The DRS220 is n...

Page 9: ...hose Open the cylinder valve and purge air from the hose 5 Program the desired refrigerant charge weight as described in the Programming section Activate the program by pressing 6 Open the manifold v...

Page 10: ...DRS220 User s Guide 16 Notes DRS220 User s Guide 17 Notes...