Introduction

The HA1 hermetic analyzer is designed to pro-
vide the service technician with an easy to use, 
reliable test instrument. To obtain the maxi-
mum benefit from 
the HA1, please take time to read these 
instructions.

Note:

 

Under some test conditions 120 V AC 

may be present on a test cable. Refer to section 
on operation.

Features include

     • Free locked rotors by reversing motor action   
     • Indicates continuity and ground faults by front 

panel indicator lights

     • Front panel input jacks to facilitate VOM 

measurement of voltage and resistance

     • Dual 120/240 V AC operation
     • Color coded for easy use

Safety Notes

Before using this instrument, read all safety  
information carefully. In this manual the word 

“WARNING”

 is used to indicate conditions or 

actions that may pose physical hazards to the 
user. The word 

“CAUTION”

 is used to indicate 

conditions or actions that may damage this 
instrument.

International Symbols

Controls and Indicators

     1.  

PWR Breaker:

 This is a combination 25A circuit 

breaker and a 

POWER ON

 switch. 

     2.  

CAP100/200/300:

 Selects one of the three starting 

capacitor ranges.

     3.  

REV/OFF/FWD:

 Switch determines the mode of 

operation of the compressor. In the 

FWD

 position, 

power is applied to the compressor motor run 
winding, and the start capacitor is connected to the 

START

 cable. In the 

REV

 position, power is applied 

to the compressor motor start winding and the start 
capacitor is connected to the 

RUN

 cable. In the 

OFF

 

position, the power circuit to the motor is broken.

The REV position is momentary. The 

OFF

 and 

FWD

 

positions are sustained.

     4.  

FAULT/CONT./START:

 Is a multi-function switch. 

The FAULT (momentary action) position is used 
to check for shorts between the COM and the RUN/
START windings. The CONT. (sustained action) posi-
tion is used to check for GROUND, RUN and START 
continuity. In the START (momentary action) position, 
the start capacitor is connected to the 
compressor motor through the REV/OFF/FWD switch.

     5.  

Ground Test (Yellow):

 This push button switch 

tests for ground continuity between system ground 
and the frame of the appliance being tested.

The HA1 introduces an “artificial ground” to make 
possible the testing for short circuits between the 
motor winding and the frame of appliances which 
may not be directly grounded. For example, units 
with two wire power cords would not be connected 
to system ground (unless a separate ground 
connection were present).

     6.  

Start Test (Red):

 This push button switch test two 

conditions, depending on the position of the FAULT/
CONT./START switch.

     7.  

Run Test (Black):

 This push button switch tests  

 

two conditions, depending on the position of the  

 

FAULT/CON./START switch.

     8.  

Indicator Light:

 LINE NORM (white)/REV (red). 

These lights are used to indicate the “condition” of 
the power source to which the HA1 is connected.

     9.  

Test Jacks:

 The four test jacks are connected 

directly to the test cable of the corresponding 
color: 

COMMON = white
GROUND* = yellow
START = red
RUN = black

The purpose of the test jacks is to facilitate 
resistance and voltage readings by enabling 
voltmeter/ohm meter test leads to be 
connected to the compressor motor circuit 
by inserting them into the appropriate jack 
at the HA1 front panel.

*Since this cable is an “artificial” ground line it is colored 
yellow instead of green.

Operating Instructions

NOTE:

 

Every effort has been made to make the 

HA1 a safe and versatile tester. However, under 
some circumstances line voltage may be present on 
the HA1 test cables when the control switches are in 
the OFF position. For this reason it is very important 
to read and become familiar with the section on 
Operating Instructions.

The following procedures have been detailed so a 
thorough understanding of the operation sequence 
may be gained. In practice, the following tests may 
be performed very quickly.

110 V AC Operation

           

CAUTION!

The HA1 is wired so that the white COMMON test 
cable is connected to the neutral side of the line. 
The red START and black RUN cables are connected 
to the hot side of the line by the POWER BREAKER 
switch and the REV/OFF/FWD switch. If the 110 V AC 
receptacle has been accidentally reverse wired, the 
white COMMON cable will be “hot” as soon as the 
HA1 is plugged into the power receptacle. To warn 
against such a condition, the red REV LINE indica-
tor will light as soon as the HA1 is plugged into a 
grounded receptacle. In this case follow procedure B 
for reversed line condition.

A.  Normal Test Procedure

    1.  The REV/OFF/FWD switch must be in the 

OFF position.

    2.  Plug the HA1 into a grounded 110 V AC 

receptacle and turn on the POWER switch. 
This is a push on - push off switch. The white 
NORM LINE indicator should light.

NOTE:

 

If the receptacle is ungrounded, both the 

NORM and the REV LINE indicator will light, but at 
half intensity. The absence of a grounded line will 
not prevent any of the following tests from being 
made. 

IMPORTANT:

  verify that the receptacle is 

not also reverse wired by measuring the voltage 
between the white COM cable and a ground 
connection. A reversed line will measure 110 V AC. 
A normal line will measure zero volts. If the 
receptacle is reverse wired, follow procedure B.

    3.  Remove power to the unit which is to be tested.

    4.  Remove and identify the connectors going to 

the common, start, and run terminals of the 
compressor motor.

    5.  Connect the test cables of the HA1 to the 

compressor motor.

       

•  Yellow GROUND cable to compressor 

motor frame

       

•  Red START cable to the start terminal

       

•  Black RUN cable to the run terminal

       

•  Leave the white COMMON cable 

disconnected until step 7

Position Range

100MFD 

88-180MFD

200MFD 

161-193MFD

300MFD 

249-301MFD

Switch Position 

Condition Tested

FAULT 

Short between START winding

     

and frame

CONT. 

Continuity of START winding

Switch Position 

Condition Tested

FAULT 

Short between RUN winding

     

and frame

CONT. 

Continuity of RUN winding

Line Voltage  Indicator Light  Status

110 V AC 

NORM on/REV off 

Normal

     

REV on/NORM off 

Neutral and line wires 

     

 

reversed at power 

     

 

receptacle. Refer to 

     

 

CAUTION under  section on 

     

 

110 V AC operation

     

NORM and REV 

System ground open or not

     

on (half intensity) 

connected at power receptacle

220 V AC 

NORM and REV 

Normal

     

on (full intensity)

HA1

Hermetic 

Compressor Analyzer