S p e c i f i c a t i o n s

Function Test

The protective cap also includes a measurement and battery verification
test function.

Measurement test:

1. Touch the electrodes to the two test points indicated with 

the “

--T--

”.

2.  Verify the display reads 27% ±2% on the wood scale.

• If it is out of this range return for repair

.

Battery test:

1. Touch the electrodes to the two test points indicated with 

the “

--B--

”.

2.  Verify the display reads 44%. 

• If the display is less than 44% replace the batteries.

M a i n t e n a n c e

These guidelines will help you attain long and reliable service from 
your meter:

1. Keep instrument dry.
2.  Keep measurement pins clean and protected.

Changing the Electrodes

One set of replacement electrodes are included in the event that the
original set is damaged or worn. To replace simply remove by hand or
use a 5mm open end wrench.

1. Remove protective cap.
2.  Remove electrodes.
3.  Install new electrodes.
4.  Replace protective cap.

WARNING!

Do not over tighten when installing the electrodes.

MM100-MAN    

P. 2

Measuring principal

Electrical resistance

Display

LCD bar graph w/dual range for wood/materials

Electrodes

Integrated and replaceable

Electrode length

5/16” (8mm)

Measuring range

Wood

6~44%

Material

0.2~2.0%

Accuracy

Wood

±1%

Material

±0.05%

Auto power off

After approx 15 minutes

Battery

(3) CR2032

Operating range

32˚~104˚F (0˚~40˚C)
0~85% RH

Dimensions

5.5” x 1.9” x 1.0” (139 x 47 x 25 mm)

Weight

0.22 lb (100g)