Power switch  

 

In center OFF position, RESP1 is powered down. 

 

Set to ON position for normal operation. 

 

Set to TEST to check battery: Battery is OK if 
meter needle is above "B/OK" line.  Otherwise, 
replace battery. 

 

Meter 

 

In normal operation, meter needle moves with 
each inspiration and expiration. 

 

 

Size control 

 

Adjusts amplitude of signal to meter and to 
downstream recorder or plotting software. 

 

Input connectors 

 

The standard RESP1 features two 0.060" safety 
jacks, black and red, that accept standard EEG 
and EKG electrode leads. 

 

The RESP1/EF and RESP1/EKG variants have an 
additional, green jack for an EKG reference lead. 

 

Output connectors 

 

The standard RESP1 has one back-panel BNC 
receptacle that supplies the respiration output 
signal to downstream equipment. 

 

The RESP1/EF and RESP1/EKG variants have an 
additional BNC receptacle to supply the EKG 
output to downstream equipment. 

 
 

 

 

Electrode considerations 

 

To improve recording quality, 

 

place electrodes at level of maximum change in 
chest diameter, usually along the mid-axillary lines 
(left and right) at about the V5 or V6 level; 

 

if signal is still weak, connect several electrodes 
together in parallel to increase effective size. 

 

If RESP1 signal is too strong for recording device, 
you will need to attenuate RESP1 output.  Contact 
UFI for a simple resistive circuit that you can use 
to decrease signal. 

 

Using the standard RESP1 

 

To measure respiration with EEG or EKG electrodes: 

 

Lightly abrade skin at the selected locations, then 
attach electrodes. 

 

Connect electrodes to front-panel safety jacks. 

 

Move power switch to TEST position.  Battery is 
OK if meter needle is above "B/OK" line (7 volts).  
Otherwise, replace battery. 

 

Move power switch to ON position.  Meter will 
swing off scale, then return to center (zero) 
position in about 30 seconds. 

 

Meter now indicates respiration; needle deflection 
magnitude reflects respiration depth. 

 

Attach recorder or other downstream device to 
respiration output connector. 

 

Output signal level ranges from 100 mV to 5 volts; 
adjust with size control. 