Page 26 of 51
Holding with Poor Sig
Select "Yes" to hold last good flow signal displayed if
the Flowmeter experiences a poor signal condition.
This function will allow to calculate flow totalizer
data without interruption. Select "NO", instead.
Empty Pipe Setup
This parameter is used to overcome the possible
problems that usually show up when the pipe being
measured is empty. Since signals can be transmitted
through the pipe wall, the flow meter may still read a
flow while measuring an empty pipe. To prevent this
condition from happening, you can specify a value.
When the signal quality falls below this value, the
measurement stops automatically. If the flow meter is
already able to stop measuring when the pipe is empty,
a value in the range of 30 to 40 should also be entered
in this window to ensure no measurement when the
pipe is empty.
Holding with PoorSig
NO
Measurement Units Options
Select the measurement unit as follows:
0. Metric
1. English
Measurement Units In
0.Metric
Flow Rate Units Options
The following flow rate units are available:
0. m3 Cubic Meters
1. 1
Liters
2.
gal
USA Gallons
3.
ig
Imperial Gallons
4.
mg
Million Gallons
5.
cf
Cubic Feet
6.
bal
USA Barrels
7.
ib
Imperial Barrels
8.
ob
Oil Barrels
The following time units are available:
/ Day
/ Hour
/ Min
/ Sec
Factory default is Cubic Meters/hour
Flow Rate Unite [31
m3/h
UFM-40
Ultrasonic transit-time flowmeter
Summary of Contents for UFM-40
Page 1: ......
Page 2: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 10: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 25: ......
Page 28: ......
Page 29: ......
Page 32: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 46: ......
Page 47: ......
Page 49: ......
Page 50: ......