Instruction for use (rev 141119)
e.ON 200 NEO / e.ON 250 NEO
UGIN DENTAIRE
17
5.2.11.
Silver test
This programme has been pre-defined and allows to calibrate the temperature of the unit in case of
unsatisfactory firing. These settings cannot be modified.
Open the firing chamber, place a 30 mm silver wire on the firing base and press the START button.
At the end of the programme, the wire should be melted only at the tip.
If not, the calibration offset may need to be modified in each programme (see chapter 5.4.1).
Repeat this process, if necessary.
5.2.12.
Decontamination
This programme is similar to the silver wire test and cannot be modified. It allows to clean the firing
chamber.
Open the furnace, place a carbon rod CARBONET on the firing base and press the START button to let the
cycle run.
5.2.13.
Auto-test
In case of a malfunction, the defective component is indicated.
ATTENTION: do not interrupt the test before completion
At the end of each phase, a logo appears indicating the result of the test.
At the end of test, the key
can be used to return to the previous screen
Exemple :
Auto-test: problem with the vacuum in the firing chamber
5.2.14.
Technician parameters