Instructions for use (rev 011221)
e.ON sinter fast +
UGIN DENTAIRE
20
6.7.3.
CLOSING THE PLATFORM AND START CYCLE
When the cycle has been selected and elements are in place
(see previous chapters):
-
close the platform: press “arrow key upward”, and start the cycle: press “RUN” key
-
or start the cycle directly with the “RUN” key
6.7.4.
SEQUENCE START OF CYCLE
a)
Start cycle “RUN”: the furnace starts to rise in temperature,
b)
the platform rises and stays closed 5 to 10 seconds,
c)
the platform goes down partially and holds about 5 min,
d)
the platform rises without closing completely,
e)
when the chamber reaches 500°C, the platform rises and closes completely.
These movements are realized
automatically
: the user does not have to intervene.
IMPORTANT
: during the platform’s movements,
DO NOT PRESS BUTTONS
(hold, stop, arrow key upward and downward):
suspend cycle
Note: when start the cycle, the noise generated by the power transformer is normal.