Instructions for use (rev 011221)
e.ON sinter fast +
UGIN DENTAIRE
16
6.6.3.
PROGRAM PRE-RECORDERED No. 3
(short cycle, for up to 5 single unit)
C01
50,0
C06
1550,0
C11
0,0
C16
0,0
T01
9
T06
15
T11
0
T16
0
C02
900,0
C07
1550,0
C12
0,0
C17
0,0
T02
6
T07
-121
T12
0
T17
0
C03
1100,0
C08
0,0
C13
0,0
C18
0,0
T03
2
T08
0
T13
0
T18
0
C04
1100,0
C09
0,0
C14
0,0
C19
0,0
T04
13
T09
0
T14
0
T19
0
C05
1400,0
C10
0,0
C15
0,0
C20
0,0
T05
14
T10
0
T15
0
T20
0
6.6.4.
PROGRAM PRE-RECORDERED No. 40 - DECONTAMINATION PROGRAM
C01
50,0
C06
900,0
C11
0,0
C16
0,0
T01
50
T06
-121
T11
0
T16
0
C02
300,0
C07
0,0
C12
0,0
C17
0,0
T02
70
T07
0
T12
0
T17
0
C03
1000,0
C08
0,0
C13
0,0
C18
0,0
T03
180
T08
0
T13
0
T18
0
C04
1620,0
C09
0,0
C14
0,0
C19
0,0
T04
240
T09
0
T14
0
T19
0
C05
1620,0
C10
0,0
C15
0,0
C20
0,0
T05
70
T10
0
T15
0
T20
0
A DECONTAMINATION PROGRAM MUST BE CARRIED OUT AS A PRECAUTIONARY MEASURE
BEFORE THE FIRST USE OF THE FURNACE
:
start the pre-recorded program number 40.
If use different types of zirconia, zirconia of different brands or zirconia with different sintering temperatures
(even during separate sintering cycles) it may be necessary to carry out
regular decontamination of the
furnace e.ON sinter fast +
using the cycle recorded in number 40.
Decontamination before the 10
th
cycle following the previous one is highly recommended, and after a long
period of non-use or at the first signs of contamination on the heating elements (whitish mark).
PRECAUTIONS:
▪
High translucency zirconia are very sensitives
. To avoid and prevent crossed discoloration with other
zirconia brand or type, user must do run more frequent decontamination programs.
▪
If
liquid shades are used
with staining technic, prosthetics units must be dry and preheated with an infra
red lamp or into an oven to prevent pollution into the e.ON sinter fast +, in following zirconia and shades
manufacturer recommendations.
Summary of Contents for e.ON sinter fast
Page 1: ...INSTRUCTIONS FOR USE Sintering furnace R v 011221...
Page 27: ......