Instructions for use (rev 011221)
e.ON sinter fast +
UGIN DENTAIRE
25
8.2.
TROUBLE-SHOOTING
Problem
Cause
Procedure to followed
The 1818.3
temperature is
displayed on the
touch screen
Defective thermocouple,
(broken or not properly
connected)
-
Do not use the furnace: risk of material
damage.
-
Turn off furnace
(switch ON/OFF).
-
Contact the after-sale service.
The result is not
satisfactory
The parameters of the
program performed
(temperature/time) are not
adequate
-
Check the program parameters: temperature
and time.
-
Adjust sintering temperatures if necessary.
(see chapter 6.4)
The furnace does
not heat up
It is necessary to test the
continuity of the rods
-
Do not use the furnace: risk of material
damage.
-
Turn off furnace (switch ON/OFF).
-
Contact the after-sale service for assistance in
the procedure.
White clusters
visible on the rods
Due to the use of different
brands or types zirconia
-
Clean the chamber
using a Nacera Clean type
product (registered trademark) and launch
several
decontamination program
(No. 40)
successive.
Zirconia pearls have
a discoloration in
yellow colour
Pearls are polluted
-
Replace
with new sintering beads.
If the failure persists, contact the after-sale service.