Instructions for use
(rev 150620)
e.ON
UGIN DENTAIRE
14
6.4.2.
PROGRAMMING INSTRUCTIONS
C: Temperature in degrees Celsius (°C)
T: Time in minutes (min)
Temperature 1
Used as the temperature start value (usually set to 50).
Time 1
Total duration of first stage (0 to 9999 minutes).
Temperature 2
Final temperature of first stage and start temperature of second stage.
Temperature values from C1 to C2 define the increase rate.
The value must be less than 10°C/min.
Time 2
Total duration of second stage (0 to 9999 minutes).
Temperature 3
Final temperature of second stage and start temperature of third stage.
If C3 is less than C2 (temperature decrease). The value must be less than 30°C/min.
Time 3
Total duration of third stage (0 to 9999 minutes).
Temperature 4
Final temperature of third stage and start temperature of fourth stage.
If necessary, the parameters are programmed as for the previous steps.
Last time
"-121"
Indicates the end of program, temperature controller works according to the
temperature/time parameters setting.
When the program detects a "-121", STOP command is activated with return to the
previously selected program.
"-121" also indicates correct cycle sequence.
ATTENTION
:
The operator must setting the temperature/time initial and the temperature/time final for
each stage
. These data must be calculated from the desired increase/decrease rate, following the
recommendations given (see examples below). Never write on the screen the desired speed directly, the
software cannot make the calculation and it takes the value as a temperature or a time “untreated”.