Instructions for use
(rev 150620)
e.ON
UGIN DENTAIRE
15
6.4.2.1.
EXAMPLES OF CALCULATION
▪
Calculation time of second stage (T02)
From the initial and final temperature values of the second stage, calculation of the time stage:
Temperature 1
→
C01 = 900°C
Temperature 2
→
C02 = 1200°C
Temperature
difference:
C02
–
C01 = 1200°C
–
900°C = 300°C
Recommended value of increase rate: less than 10°C/minute
300°C at a rate of 10°C/min
→
300°C / 10°C = 30 min
→
T02 = 30 minutes
▪
Parameters for remain to end temperature (T04)
Parameters temperature/time of end stage to obtain remains to the end temperature of 1520°C for 2 hours:
Temperature 4
→
C04 = 1250°C
Temperature 5
→
C05 = 1250°C
Temperature difference = 0°C:
Temperature remains to 1250°C
Time remains: 2 hours
Time 04:
2 hours x 60 min = 120 min
→
T04 = 120 minutes
Zirconia
SINTERING TEMPERATURES
are given as indicative by disc manufacturer. If necessary,
these one
can be adjusted from -15°C to +15°C
around the final temperature recommended.
We recommend adjusting the temperature/time for special work.
6.5.
PARAMETER SETTING
This interface requires a password to be accessible.
No data may be modified without authorization:
risk of
damages on the furnace or trigger heating untimely.
The modification of the parameters is therefore
reserved
for technicians of customer’s service
UGIN DENTAIRE.
NOTE: even if changes are made to the settings, when saving them, a screen is displayed to insert the
password. When no password has been inserted and validation is done with the "Ok" key, the changes made