Instructions for use
(rev 150620)
e.ON
UGIN DENTAIRE
18
6.7.
POSITIONING THE ELEMENTS AND STARTING THE CYCLE
6.7.1.
POSITIONING OF SUPPORT
▪
For use the furnace,
always place the support on the platform,
including for the decontamination
program.
▪
Use only the original support
e.ON
UGIN DENTAIRE.
▪
Before use,
check if it is in perfect condition
: presence of dirt or damage.
Clean with brush if necessary
(see chapter 8).
If the support is damaged or cracked, do not use and replace it.
▪
It is composed of a base and a floor.
According to the number of tray for sinterizing to position, it is necessary to place or remove the floor:
6.7.2.
POSITIONING OF ELEMENTS AND TRAY FOR SINTERIZING
▪
Empty a zirconia pearls
into a tray (a layer at the bottom to prevent the elements from rubbing on the
bottom and sides).
▪
Place the elements in a tray
(avoid touch amongst themselves).
▪
The three trays can be stacked
: you can stack 2 or 3 trays.
▪
Always place the cover on the last tray
(only on the last, never between two trays).
▪
Place the tray(s) at the centre of support.
NUMBER OF TRAYS
ORDER
1
Tray No. 1 + cover
2
Tray No. 1 + Tray No. 2 + cover
3
Tray No. 1 + Tray No. 2 + Tray No. 3 + cover
-
Capacity till 60 units.
-
Beads for sinterizing must be changed in any tray if a discoloration in yellow colour appears.
NUMBER OF TRAY
TO PLACE
SUPPORT TO PLACE
1
base + floor
2
base + floor
3
base
IMPORTANT: observe these indications so as not to
exceed the maximum height (risk of damage)