Instructions for use
(rev 150620)
e.ON
UGIN DENTAIRE
22
SUPPORT
(see chapter 6.7.1
–
6.7.2)
-
Place the support on the platform
while the
furnace is working.
-
Observe the information on the support:
base
and floor
(chapter 6.7).
-
Always place the elements in the tray for
sinterizing, with the zirconia beads inside, and
place the cover
on the last tray
(never between
two trays).
-
Beads for sintering must be changed in any tray
if a
discoloration in yellow colour
appears.
DECONTAMINATION PROGRAM
(see chapter 6.6.1)
-
Before first sintering, user must launch a
decontamination program (No. 40).
-
Always place the support (base + floor) on the
platform during carry out a decontamination
program.
-
Add pieces of white Zirconia (absorption of
pollutants) to improve the
effectiveness of
decontamination
:
-
yellow colour = enclosure still
contaminated: repeat a decontamination
cycle.
-
black colour = ok: furnace ready.
-
High translucency zirconia
are very sensitives
(crossed discoloration risk): user must do
regularly decontamination program.
-
If
liquid shades
are used with staining technic,
prosthetics units must be dry and preheated
with an infra-red lamp or into an oven to
prevent pollution into the
eON SINTER
+, in
following zirconia and shades manufacturer
recommendations.
-
White clusters can appear on the rods
(heating
elements) when different brands or types of
zirconia are used: user must clean chamber and
rods with a special zirconia cleaner like
Nacera
Clean (registered brand)
with decontamination
program.
Note: observe the instructions set out in chapter 8
–
maintenance.