Instructions for use
(rev 150620)
e.ON
UGIN DENTAIRE
16
6.6.
PROGRAMS
6.6.1.
PROGRAMS PRE-RECORDERED: No. 1 and No. 40 (decontamination)
Two programs are pre-recorded in the furnace:
PROGRAM No. 1
DECONTAMINATION PROGRAM
–
No. 40
STAGE
TEMPERATURE (°C) TIME (min)
STAGE
TEMPERATURE (°C)
TIME (min)
1
C1: 50
T1: 25
1
C1: 50
T1: 40
2
C2: 300
T2: 40
2
C2: 300
T2: 70
3
C3: 600
T3: 40
3
C3: 1000
T3: 150
4
C4: 1000
T4: 20
4
C4: 1620
T4: 60
5
C5: 1200
T5: 10
5
C5: 1620
T5: 70
6
C6: 1200
T6: 25
6
C6: 900
T6: -121
7
C7: 1450
T7: 30
8
C8: 1450
T8: 30
9
C9: 1200
T9: 30
10
C10: 1200
T10: -121
A DECONTAMINATION PROGRAM MUST BE CARRIED OUT AS A PRECAUTIONARY MEASURE
BEFORE THE FIRST USE OF THE FURNACE
: start the pre-recorded program number 40.
If use different types of zirconia, zirconia of different brands or zirconia with different sintering temperatures
(even during separate sintering cycles) it may be necessary to carry out regular decontamination of the
furnace e.ON using the cycle recorded in number 40.
PRECAUTIONS:
▪
High translucency zirconia are very sensitives
. To avoid and prevent crossed discoloration with other
zirconia brand or type, user must do regularly decontamination program.
▪
If
liquid shades are used
with staining technic, prosthetics units must be dry and preheated with an infra
red lamp or into an oven to prevent pollution into the eON , in following zirconia and shades
manufacturer recommendations.
▪
White clusters can appear on the rods
(heating elements) when different brands or types of zirconia are
used. In this case user must clean muffle and rods with a special zirconia cleaner like Nacera Clean
(registered brand) with decontamination program No. 40.
▪
Always place the support (base + floor) on the platform
during carry out a decontamination cycle.
ADVICE: add pieces of white Zirconia
(absorption of pollutants)
to improve the effectiveness of
decontamination:
-
yellow colour = enclosure still contaminated: repeat a decontamination cycle
-
black colour = ok: furnace ready