INSTRUCTIONS FOT USE 

 

LECTRA PRESS 

27 

3  

CALIBRATION INDICATOR

 

 
All  Ugin’Dentaire  furnaces  have  been  subjected  to  IPTS  68  (  International  Practical 
Temperatue  Scale,  1968  )  temperature  testing  using  a  Eurotherm  239-type  millivolt 
generator. 
You can, however, modify the furnace settings by means of digital calibration. 
This easy-to-use function applies to all furnace programs. 
When programming the final temperature, press the standby key Standby
The calibration indicator lights up and the parameters display shows the factory preset 
calibration number, 

5

This setting assures satisfactory operation for a new muffle. 
 
If the firing results indicate an overly high baking temperature 
 
 

 

 

 

- increase the setting :  

6

 = 1 point, or : -10°c 

 

 

 

 

 

 

 

 

7

 = 2 points, or : -20°c 

if the firing results indicate an overly low temperature 
 
 

 

 

 

- decrease the setting : 

4

 = 1 point, or : + 10°c 

 

 

 

 

 

 

 

 

3

 = 2 points, or : +20°c 

You can check the accuracy of the temperature setting by running a silver test. 
( see page 31 ) 
 

4  

PROGRAM NUMBER DISPLAY

 

 

The programming unit can memorize up to 100 programs ( 00 to 99 ). 
N° 00 to 89 : porcelain firing programs 
N° 90 to 99 : pressing programs
 
To  call  up  a  program,  press  clear  key  CL  and  then  enter  the  program  number  on  the 
digital keyboard. 
This operation can only be carried out if the furnace is in standby mode ( no display on 
diagram 1 ) and the platform is in low position. 
The  furnace  is  in  standby  mode  when  the  furnace  is  switched  on,  after  pressing  the 
Stop key or standby key Standby ( if need be ) or at the end of a firing cycle. 
 

 FIRING PARAMETER DISPLAY 

 
This display gives the temperature of the program called up. 
When programming, it displays the values of the various parameters of a program. 
To move from one parameter to the next, press the 

 key. 

 
 
 

Summary of Contents for LECTRA PRESS

Page 1: ...DESCRIPTION OF REAR PANEL 24 DESCRIPTION OF THE PROGRAMMING UNIT 25 to 30 PARAMETERS OF PORCELAIN FIRING 30 31 32 33 PARAMETERS OF A PRESSING PROGRAM 34 OPERATION OF A PRESSING PROGRAM 34 35 MAINTENA...

Page 2: ...ss Heat rise rate 0 200 c min Vacuum start 0 1000 c Vacuum level 0 99 Prog Ceram 50 99 Prog Press Vacuum stop 0 1200 c Prog Ceram 1250 c Prog Press Final temperature 0 1200 c Hold time 0 60 min Intern...

Page 3: ...m pump socket 4 Plug the vacuum hose into the connector 5 Plug the socket 3 Plug the air pressure hose into the connector 7 WARNING For the furnace to work properly the circuit air pressure should obl...

Page 4: ...L 9 5 10 1 2 3 4 6 7 8 1 Mains switch on off 2 Thermal circuit breaker 3 Socket 4 Socket for connection of vacuum pump 5 Connector for vacuum hose 6 Name plate 7 Connector of pressure 8 Pressure regul...

Page 5: ...ndicator 4 Program number display 5 Firing parameters display 6 Indicator of units display on 5 7 Vacuum level display 8 Temperature display 9 Time display 10 Start key begins a cycle 11 Stop key 12 P...

Page 6: ...has been selected the firing cycle begins The indicators corresponding to all the operating stages light up at once As each stage in the program is completed the corresponding light goes out Note you...

Page 7: ...ts indicate an overly low temperature decrease the setting 4 1 point or 10 c 3 2 points or 20 c You can check the accuracy of the temperature setting by running a silver test see page 31 4 PROGRAM NUM...

Page 8: ...tes the countdown for the various stages 10 START KEY Starting the cycle After checking that you have called up the appropriate program number press the Start key to begin the firing cycle All the ind...

Page 9: ...matic standby If the furnace is not operated for five minutes after the end of a firing cycle the platform lifts and the furnace switches to standby position 300 c furnace closed To exit from the stan...

Page 10: ...ring the entire platform lift time PLATFORM LIFT TIME 0 TO 30 min The platform lift time is divided into six stages The platform only lifts once the pre drying temperature has been reached While the p...

Page 11: ...ncel the vacuum or maintain it starting at a specified temperature VACUUM HOLD TIME 0 to 30 min At the vacuum stop temperature you can program the vacuum to continue for a length of time between 0 and...

Page 12: ...to 30 minutes Otherwise to open the platform immediately program INTERNAL COOLING 0 EXTERNAL COOLING TIME 0 to 30 min The platform descends in six stages Once the platform has reached its lowest posi...

Page 13: ...0 c 450 c PLATFORM RAISED 0 to 30 min 0 INTERNAL PRE DRYING 0 to 900 c 700 c HEATING RATE 0 to 200 c min 60 c min VACUUM START 0 to 1000 c 700 c VACUUM LEVEL 50 to 99 90 VACUUM STOP 1250 c unchangeabl...

Page 14: ...with two centred diameters has been placed on the furnace platform A program is initiated by pressing the Start key once unless the appliance is in the standby position platform raised DESCRIPTION OF...

Page 15: ...times 17 Press the Start key 18 The platform lowers 19 End of cycle NB It is recommended that a cylinder clamp be provided close to the machine IMPORTANT While the cycle is running you can press the S...

Page 16: ...he panel with a solvent free product Keep hot pieces away from the panel to avoid the risk of damage by heat radiation Leave the furnace hooked up all the times except during prolonged absence DECONTA...

Page 17: ...ery high over voltage may lead to errors in program data To guard against such an eventuality it is wise to note your usual programs in the booklet supplied with the furnace POWER FAILURE OR VOLTAGE F...

Page 18: ...seconds All 11 led will blink on and off for 5 seconds after which time only Led pre drying temperature stays lit List of parameters WARNING All other parameter modification can lead to important pro...

Page 19: ...sts contact your after sales service dF Plab Malfunction platform in low position press P to stop the buzzer then check that no object is preventing the platform lowering If so turn off switch 1 to cu...

Page 20: ...rature 600 c Dry time 0 Internal pre drying 0 Heat rise 50 c min Vacuum start 0 Vacuum level 0 Vacuum release hold 0 High temperature 960 c Hold time 1 30 min Internal cooling 0 Lowering platform 0 If...

Reviews: