Instructions for use (rev 031121) - En
PROGRAMIX 50 - 100
UGIN DENTAIRE
11
5.2.
OPERATING PARAMETERS
-
To program the operating parameters, press key
-
When the program number blinking (display No.2), enter the data with keys
and
-
To switch to another parameter, use the
key.
-
When you have programmed all the parameter, “lock” the pr
ogram by pressing key
again.
5.3.
DEPARTURE DELAY
This timer function is used to delay the beginning of the firing cycle:
from 0 hours to 99 hours and 59 minutes.
A count-down before start time is displayed on the control panel (n° 3).
Should you not wish to use this function, set the parameter at 0h 0min, heating will begin immediately (or
use the
key after pressing the
touch control).
5.4.
1
st
, 2
nd
AND 3
rd
STAGE AND DELAY FINAL
▪
1st STAGE
o
Rate No. 1: STOP
–
from 1 to 15°C/min - FULL
The heating rate is the heat rise per unit time (rate of rise). There are three positions:
-
STOP
(the furnace stops heating)
: this is only used when the next stage is at a Iower temperature.
-
1 to 15°C/min
:
h
eating is regulated and will not rise above the programmed value. However, it may
be lower than that value, especially in the higher temperature ranges, or if the muffle is weIl Ioaded
(several cylinders inside the muffle).
-
''FULL''
: this position corresponds to the maximum furnace output: the heat rise rate is no longer
regulated.
o
1
st
stage temperature: from 0 to 1200°C
Temperature setting for the 1
st
stage.
o
Hold time from 0 to 240 min
Hold time for the temperature of the 1
st
stage.
▪
2
nd
STAGE
o
Rate No. 2 (RP 2): from 1 to 15°C/min
–
FULL - STOP
Heat rise to reach the second stage. The principle is the same as for heating rate No. 1, except that the
programmed values may differ.
o
2
nd
stage temperature: from 0 to 1200°C
Temperature setting for the 2
nd
stage.
o
Hold time from 0 to 240 min
Hold time for the temperature of the 2
nd
stage.