UGO BASILE - I
TALY
Page 11
Series 58000 Instruction Manual (Rev.1)
PULSE TRANSDUCER:
for connection of the selected transducer, see paragraph 2.5.
PNEUMATIC
CONNECTION:
for connection of the selected cuff, see paragraph 2.4.
4.5
Additional Safety Consideration
a.
Place your BP Recorder on a steady flat surface.
b.
Do not obstruct free and comfortable access to the power module.
c.
Use original accessories and spare parts only, see also paragraph 13-
ORDERING INFORMATION.
d.
Immediately disconnect and replace an accidentally damaged mains cable.
e.
Do not operate the BP Recorder in hazardous environments or outside pre-
scribed environmental limitations (i.e. 10°C / +40°C, 95% relative humidity,
non-condensing).
f.
Do not spray any liquid on the connectors.
UGO BASILE DOES NOT ACCEPT ANY RESPONSIBILITY FOR PROB-
LEMS OR HARM CAUSED TO THINGS OR PERSONS ARISING FROM:
♦
INCORRECT ELECTRICAL SUPPLY.
♦
INCORRECT INSTALLATION PROCEDURE.
♦
INCORRECT OR IMPROPER USE OR, IN ANY CASE, NOT IN AC-
CORDANCE WITH THE PURPOSE FOR WHICH THE INSTRUMENT
HAS BEEN DESIGNED AND THE WARNINGS BROUGHT TO YOUR
ATTENTION IN THE INSTRUCTION MANUAL SUPPLIED WITH THE
INSTRUMENT.
♦
REPLACEMENT OF ORIGINAL COMPONENTS, ACCESSORIES OR
PARTS WITH OTHERS NOT APPROVED BY THE MANUFACTURER.
♦
SERVICING CARRIED OUT BY UNAUTHORIZED PERSONNEL.
5
PRELIMINARY
5.1
Pre-warming the Animal (Rat & Mouse)
Place the animal in a warming cupboard, at about 29-30°C for at least 30 minutes, in order to
cause a sufficient vasodilation in the caudal artery. We recommend our models, purposefully de-
signed for this task, see 2.7-Optional Animal Heater /Scanner.
Some researchers feel that in this way the vasodilation of the whole vascular system will inevita-
bly cause a “pressure drop”. This pressure drop takes place in all animals and consequently the
reliability of a pressure screening should not be severely affected. Admittedly, this reasoning may
leave some scientists uncomfortable about the artefact he/she feels it may impair the experiment.