UGO BASILE - I
TALY
Page 17
Series 58000 Instruction Manual (Rev.1)
At this point, select among automatic (
F1 AUT
) or manual (
F2
MAN
) measurement.
7.1
Automatic
Measurement
The instrument takes about three seconds to check the pulse
amplitude, see the picture on the right.
The amplifier gain is then automatically adjusted, in order to
maintain the amplitude of the graph within convenient ampli-
tude.
In the following two seconds the average heart rate, in pulses
per minute, is calculated.
If the signal is sufficient
, the following display will appear:
note the caption “
RELIABLE SIGNAL”.
Depress the blue key
START/REVERSE
, to begin the meas-
uring sequence, see paragraph 7.3-Starting the Pressure
Measurement.
A bar proportional to the amplitude of the signal, appears
above the bottom line, to monitor the signal in real time.
In the case the
signal amplitude is not sufficient
, the fol-
lowing display will appear: note the caption “
ERRATIC SIG-
NAL, REPEAT TEST””.
Depress
F4
(
ESC
). From the menu which appears, depress
F1
(
AUT
) if you decide to repeat the measure automatically.
If you opt for a manual measure, depress
F2
(
MAN
) and see
paragraph 7.2-Manual Measurement. For reference
7.2
Manual
Measurement
If the manual measurement has been selected, see paragraph 7-TEST MENU, the signal is dis-
played as one of the two following examples: