3™
Assembly and disassembly instructions
4
Special cleaning instructions
Manual cleaning
Difficult to reach parts must be
pre-cleaned manually with a brush
in a water bath. The movable sleeve
and the locking mechanism with a
push-button must be moved back
and forth and brushed during the
manual cleaning in the water bath.
The instrument must then be rinsed
thoroughly, to avoid any residue
from drying.
!
Important:
All lumens and
cavities must be rinsed out carefully
and thoroughly.
Visual inspection
Finally, it must be checked whether
there is any visible soiling. If this is
the case, the previous steps must
be repeated.
Maintaining the screwdriver
For their maintenance, movable
parts such as joints, hinges and
threaded parts should be treated
with suitable maintenance oil.
In addition, the general cleaning
instructions of the Processing
manual implants and instruments
UH 1100 also apply.