neon
3™
Assembly and disassembly instructions
6
Special cleaning instructions
Cleaning the handle
Difficult to reach parts should
be pre-cleaned manually with a
brush in a water bath. Springs
and mechanisms must be moved
back and forth several times. The
instrument must then be rinsed
thoroughly, to avoid any residue
from drying.
In addition, the general cleaning
instructions of the Processing
manual implants and instruments
UH 1100 also apply.
Cleaning the outer sleeve
Difficult to reach parts should be
pre-cleaned manually with a brush
in a water bath. In addition, the
notch at the end of the outer sleeve
of the rod reduction instrument
should be brushed. The instrument
must then be rinsed thoroughly, to
avoid any residue from drying.
In addition, the general cleaning
instructions of the Processing
manual implants and instruments
UH 1100 also apply.