neon
3™
Assembly and disassembly instructions
6
neon
3™
Assembly and disassembly instructions
for CS 3938-07, -08 and -09
Cleaning the inner sleeve
Difficult to reach parts should be
pre-cleaned manually with a brush
in a water bath. The instrument
must then be rinsed thoroughly, to
avoid any residue from drying.
In addition, the general cleaning
instructions of the Processing
manual implants and instruments
UH 1100 also apply.
Visual inspection
Finally, it must be checked whether
there is any visible soiling. If this is
the case, the previous steps must
be repeated.
In addition, the general cleaning
instructions of the Processing
manual implants and instruments
UH 1100 also apply.
Components maintenance
For the maintenance of the
components, movable parts,
threads, hinges and buttons
should be treated with suitable
maintenance oil.
In addition, the general cleaning
instructions of the Processing
manual implants and instruments
UH 1100 also apply.