Page 40
LED status messages
LAS 260.1
– Date 07/2023
Status red
NOTICE
NOTICE
Critical message
Flashing:
The unit is unable to detect volumetric flow across the filters.
It is possible that no filter is installed in the unit. The particulate
filters are not correctly seated in the unit. The intake is
possibly blocked. The set volumetric flow rate is too low. The
blower is possibly defective.
•
Insert a particulate filter.
•
Check the particulate filter for damage and correct
seating.
•
Check the intake section for free passage.
•
Increase the set volumetric flow rate.
•
Have the blower checked.
Continuously on:
The particulate filters are completely saturated.
Carry out a particulate filter change as described in the
section “Maintenance and servicing.”
Summary of Contents for LAS260.1
Page 2: ...ULT AG Am G pelteich 1 02708 L bau Germany Phone 49 3585 4128 0 Fax 49 3585 4128 11...
Page 45: ......
Page 46: ......
Page 48: ......
Page 50: ......