Page
23
of
40
Issue: 6: 08/11/2018
DRAFT
13.1.2
In line Grab Rail
1. Fit the in line grab rail as per the instructions above.
2. If using the In-line grab rail in conjunction with side rails ensure a full risk assessment
has been undertaken as this will generate entrapment risks. Using the In-line Grab rail in
conjunction with side rails must be agreed with a clinician.
13.1.3
Daily Checks on Grab Rail
Caution! Risk of entrapment
The daily maintenance work can be performed by the nursing personnel or carer:
Check the gap between the bed ends and the grab rail are less than 60mm or
greater than 318mm.
Check the thumb wheel fastenings are secure.
13.2 Attaching Side Rails to the bed
When attaching side rails the consideration of safety is paramount. Side rails that have been
manufactured by Ultimate Healthcare carrying the product codes UBEB7002-2 have been
specifically designed for the Cura range of beds; however you must ensure the following points
are adhered to:
1. Position the Side Rails according to the labels on the bed and side rails. Tilt the Side Rail
forward and clip over the bed platform, then slide the bottom fixing under the bed
platform by straightening up the Side Rail.
2. Ensure that the rails are securely fitted before use. Check the security of the fittings
every day.
3. The gap between the side rail and the bed end must be less than 60mm or more than
318mm to minimise the risk of entrapment and to comply with entrapment guidance and
regulations.
4. Ensure that the distance from the top of the uncompressed mattress and the top of
the side rail is at least 220mm.