page 22
remove air. Allow pump to run for five to ten minutes to release any air from the
system.
b. If the filter is free of air, run a few sample lenses to ensure there are no streaks
or defects.
OPTIONAL:
WARNING!
Allow the machine to cool for 30 - 60 minutes before flushing with
acetone!
It may be necessary to flush the coating pump with acetone if your coating has been
contaminated. With the tubing disconnected from the bottom of the filter, pour acetone into the
machine as you would with coating. DO NOT run acetone through the coating filter, simply flush
into a waste container.
Changing Filters
Hepa Pre-filter
1. Replace with new filters monthly
Hepa Filter
1. Disconnect power cable from filter unit to the top of the machine
2. Remove filter unit from machine
3. Remove 4 self-tapping screws holding filter onto filter unit
4. Place sealing foam around top edge of new HEPA filter
5. Place filter unit on top of new HEPA filter
6. Screw in self-tapping screws through filter unit into new HEPA filter
7. Place filter unit into machine
8. Connect power cord to top of machine
Replace Air Filter Packs
1. Turn air to the machine off
2. Press up on the bottom of the air filter canister and turn clockwise
3. Unscrew plastic retainer at the base of filter and remove old filter
4.
Install new filters marker “DX” “BX” “000” in this order
5. Screw on plastic retainer
6. Reattach air canister pressing up and turning counter clockwise
Replacing Consumable Parts
Light Module
1. Select MACHINE SET UP
2. Select REMOVE LAMP